Sedline® Miscalculation of Depth of Anaesthesia Variables in Two Pigs Due to Electrocardiographic Signal Contamination
Open Access
Abstract
Two young (11-week-old) pigs underwent sole propofol anaesthesia as part of an experimental study. The depth of anaesthesia was evaluated both clinically and using the electroencephalography(EEG)-based monitor Sedline; in particular, the patient state index, suppression ratio, raw EEG traces, and its spectrogram were assessed. Physiological parameters and electrocardiographic activity were continuously monitored. In one pig (Case 1), during the administration of high doses of propofol, the Sedline-generated variables suddenly indicated an increased EEG activity while this was not confirmed by observation of either the raw EEG or its spectrogram. In the second pig (Case 2), a similar event was recorded during euthanasia with systemic pentobarbital. Both events happened while the EEG activity was isoelectric except for signal interferences and synchronous in rhythm and shape with the electrocardiographic activity. The suggestion of increased brain activity based on the interpretation of the Sedline variables was suspected wrong; most probably due to electrocardiographic interferences. In pigs, the patient state index and suppression ratio, as calculated by the Sedline monitor, could be influenced by the electrocardiographic activity contaminating the EEG trace, especially during otherwise isoelectric periods (strong EEG depression). Visual interpretation of the raw EEG and of the spectrogram remains necessary to identify such artefacts.Keywords
Funding Information
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung (CRSK-3_190926/1)
- Berne University Research Foundation and BEKB Förderfonds (CRSK-3_190926/1)
This publication has 19 references indexed in Scilit:
- Clinical Electroencephalography for AnesthesiologistsAnesthesiology, 2015
- Comparison of Heart Rate and Blood Pressure with Toe Pinch and Bispectral Index for Monitoring the Depth of Anesthesia in Piglets.2015
- Comparison of ketamine–dexmedetomidine–methadone and tiletamine–zolazepam–methadone combinations for short-term anaesthesia in domestic pigsThe Veterinary Journal, 2015
- A new method for removal of powerline interference in ECG and EEG recordingsComputers and Electrical Engineering, 2015
- ECG contamination of EEG signals: effect on entropyJournal of Clinical Monitoring and Computing, 2015
- Assessment of consciousness during propofol anaesthesia in pigsVeterinary Record, 2011
- Removal of ECG artifacts from EEG using a modified independent component analysis approach2008
- Quantification of Epileptiform Electroencephalographic Activity during Sevoflurane Mask InductionAnesthesiology, 2007
- Bispectral index, spectral edge frequency 95% and median frequency recorded at varying desflurane concentrations in pigsResearch in Veterinary Science, 2006
- EMG contamination of EEG: spectral and topographical characteristicsClinical Neurophysiology, 2003