The purposes of this study were to (1) validate the accuracy of cine phase-contrast MR flow measurements within the portal vein, (2) develop a suitable protocol for using this method to measure volumetric flow rate in the portal venous system, and (3) use this protocol, with Doppler sonography as a reference, to measure portal venous flow in healthy volunteers and in patients with protal venous hypertension.Flow determinations were obtained in a model of fluid movement approximating blood-flow conditions in the portal venous system. A suitable protocol was based on consideration of the theoretical effects of (1) spatial resolution, (2) obliquity of the imaging plane to the direction of flow, and (3) signal-to-noise ratio of the signed quantitative velocity images (in three volunteers) on the accuracy and precision of flow measurements. This protocol was used to obtain cine phase-contrast MR images of the portal venous system in five volunteers and six patients.Values obtained with a flow phantom showed go...