Measurement of central augmentation index by three different methods and techniques: Agreement among Arteriograph, Complior, and Mobil‐O‐Graph devices
Open Access
- 29 August 2019
- journal article
- research article
- Published by Wiley in The Journal of Clinical Hypertension
- Vol. 21 (9) , 1386-1392
Abstract
Wave reflection at central arteries consists of a major component of left ventricular afterload. Central augmentation index (AIx) is the most widely used surrogate of wave reflection. Recent technological developments now provide the ability to obtain, non‐invasively, aortic, or carotid pressure waves and measure AIx based on various algorithms of pulse wave analysis. The aim of this study was to compare AIx measurements performed by the Arteriograph, Complior, and Mobil‐O‐Graph apparatuses. Recordings by each device in randomized order were performed with 5‐minute interval at 211 individuals (age 55.1 ± 14.1 years, 67.8% males) who underwent diagnostic cardiovascular assessment. All measurements were obtained at the supine position, and AIx was calculated using the formula AIx = 100 × (Augmentation pressure)/(Pulse Pressure). Bland‐Altman analysis was performed. Mean difference (bias) ± one standard deviation of difference (with limits of agreement) of AIx between different devices was as follows: (a) Mobil‐O‐Graph vs Complior: −2.1 ± 14.8% (−31.1% to 26.9%), (b) Arteriograph vs Complior: 12.9 ± 14.6% (−15.7% to 41.5%), and (c) Mobil‐O‐Graph vs Arteriograph: −10.8 ± 16.9% (−43.9% to 22.3%). The three examined devices exerted significant differences in central AIx estimation which makes the three devices non‐interchangeable for wave reflection assessment. However, the Mobil‐O‐Graph device showed the highest agreement (lowest bias) with the Complior system as regards to the AIx measurement.Keywords
This publication has 20 references indexed in Scilit:
- Pulse wave analysis using the Mobil-O-Graph, Arteriograph and Complior device: a comparative studyBlood Pressure, 2019
- Validation of Central and Peripheral Non-Invasive Hemodynamic Variables Using an Oscillometric MethodHigh Blood Pressure & Cardiovascular Prevention, 2017
- Sex differences in aortic augmentation index in adolescentsJournal Of Hypertension, 2017
- A new oscillometric method for pulse wave analysis: comparison with a common tonometric methodJournal of Human Hypertension, 2010
- Non-Invasive Methods and Techniques for Central Blood Pressure Estimation: Procedures, Validation, Reproducibility and LimitationsCurrent Pharmaceutical Design, 2009
- Changes in Aortic Stiffness and Augmentation Index After Acute Converting Enzyme or Vasopeptidase InhibitionHypertension, 2005
- Pulse wave analysisBritish Journal of Clinical Pharmacology, 2001
- Genesis of the normal and abnormal arterial pulseCurrent Problems in Cardiology, 2000
- Noninvasive determination of age-related changes in the human arterial pulse.Circulation, 1989
- Aortic input impedance in normal man: relationship to pressure wave forms.Circulation, 1980