The Practicalities of Obtaining and Using Hydrophone Calibration Data to Derive Pressure Waveforms
- 27 July 2016
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 64 (1) , 126-140
Abstract
This paper considers the means by which calibration data are used to convert hydrophone output voltage into pressure. Hydrophone frequency responses are complex-valued quantities, and only by correcting for the magnitude and phase variations, is it possible to accurately recover the original pressure waveform. The limitations of current hydrophone calibration techniques are discussed, and a new method of obtaining hydrophone phase data is presented. Magnitude and phase information is measured via both coarse increment (1 MHz) and fine increment (50 kHz) calibration techniques for three exemplar hydrophones (0.5 mm needle, 0.2 mm needle, and 0.4 mm membrane). Frequently hydrophone calibration data are available at frequency increments that do not match that required by the deconvolution process. Therefore, a variety of methods to interpolate the calibrated system response to obtain correctly spaced data are considered, and two spline interpolation methods are found to offer best performance. Data preconditioning and filtering to address artifacts above and below the 1 to 40 MHz bandwidth of the coarse frequency increment calibration are also investigated, and a simple procedure for selecting an appropriate low-pass filter is presented. The revised calibration data are used to deconvolve the hydrophone frequency response for experimentally derived waveforms. Standard ultrasonic output parameters (such as peak compressional and peak rarefactional pressures, pulse intensity integral, and temporal peak and pulse average acoustic intensities) are calculated from these waveforms. Although the three hydrophones used in this paper are of different types and have a range of active element sizes, all output parameters derived from the deconvolved waveforms have <;5% variation from their respective population means (with the majority being within <;2%).Keywords
Funding Information
- U.K. Department for Business, Innovation and Skills through the Acoustics and Ionising Radiation Metrology Programme
This publication has 40 references indexed in Scilit:
- Robust spot-poled membrane hydrophones for measurement of large amplitude pressure waveforms generated by high intensity therapeutic ultrasonic transducersThe Journal of the Acoustical Society of America, 2016
- Comparison of hydrophone phase response obtained via time delay spectrometry measurement and Hilbert transformationAIP Conference Proceedings, 2012
- 1-3 piezoelectric composite transducers for swept-frequency calibration of hydrophones from 100 kHz to 2 MHzThe Journal of the Acoustical Society of America, 2004
- Voltage to pressure conversion: are you getting `phased' by the problem?Journal of Physics: Conference Series, 2004
- Improvement of hydrophone measurements on diagnostic ultrasound machines using broadband complex-valued calibration dataJournal of Physics: Conference Series, 2004
- Sensitivity measurements of piezoelectric polymer hydrophones from 0.2–2 MHz using a broadband-pulse techniqueThe Journal of the Acoustical Society of America, 1999
- Suppression of Irregularity in Frequency Response and Directivity of a Piezoelectric Polymer Film HydrophoneJapanese Journal of Applied Physics, 1994
- Frequency response of PVDF needle-type hydrophonesUltrasound in Medicine & Biology, 1994
- The influence of waveform distortion on hydrophone spatial-averaging corrections—Theory and measurementThe Journal of the Acoustical Society of America, 1992
- The scattering of a scalar wave by a semi-infinite rod of circular cross sectionPhilosophical Transactions of the Royal Society of London. Series A, Mathematical and Physical Sciences, 1955