Abstract
OBJECTIVES The aim of this study was to investigate the mid-term haemodynamic and clinical results after aortic valve replacement (AVR) using the Sorin Freedom Solo (SFS) stentless bioprosthesis, compared with the standard Carpentier Edwards Perimount (CEP) stented bioprosthesis. METHODS In this retrospective cohort study of prospectively collected data, 116 patients were included in the SFS group (53 males; median age 74 years, range 56–85 years), and 122 patients in the CEP group (85 males; median age 73 years, range 43–88 years) between July 2007 and January 2013. Echocardiography was performed at 6 weeks after surgery in our centre, and the most recent echocardiography (in our centre or in referring cardiology departments) was requested. Between September 2013 and April 2014, all patients were called by the same researcher to gain clinical follow-up data. RESULTS Mid-term mortality was 16.4% in the SFS group (19 patients) and 21.3% in the CEP group (26 patients); (P = 0.3). The mean transvalvular gradient was 7.4 ± 3.1 mmHg in the SFS group, and 11.6 ± 3.2 mmHg in the CEP group at 6 weeks postoperatively (P < 0.001). When stratified by labelled valve size, mean gradients were significantly lower in the SFS group for every size (P ≤ 0.03). After 3.3 ± 1.4 years of follow-up, the mean gradient was still significantly lower in the SFS group than that in the CEP group (P < 0.001). Clinical follow-up showed relatively low complication rates. CONCLUSION These data suggest that the Sorin Freedom Solo stentless bioprosthesis is as safe as the Carpentier Edwards bioprosthesis, and provides better short- and mid-term haemodynamic performance than the Carpentier Edwards bioprosthesis.Keywords
This publication has 21 references indexed in Scilit:
- Single-center experience using the Freedom SOLO aortic bioprosthesisThe Journal of Thoracic and Cardiovascular Surgery, 2013
- Early clinical and haemodynamic results after aortic valve replacement with the Freedom SOLO bioprosthesis (experience of Italian multicenter study)European Journal of Cardio-Thoracic Surgery, 2012
- Aortic valve surgery: Marked increases in volume and significant decreases in mechanical valve use—an analysis of 41,227 patients over 5 years from the Society for Cardiothoracic Surgery in Great Britain and Ireland National databaseThe Journal of Thoracic and Cardiovascular Surgery, 2011
- Choice of Prosthetic Heart Valve in AdultsJACC, 2010
- The Sorin Freedom SOLO stentless aortic valve: Technique of implantation and operative results in 109 patientsThe Journal of Thoracic and Cardiovascular Surgery, 2010
- Engage People, Retire ThingsInnovations: Technology, Governance, Globalization, 2009
- Hemodynamic Performance of Stentless Versus Stented Valves: A Systematic Review and Meta-AnalysisJournal of Cardiac Surgery, 2008
- Meta-Analysis of Valve Hemodynamics and Left Ventricular Mass Regression for Stentless Versus Stented Aortic ValvesThe Annals of Thoracic Surgery, 2007
- Recommendations for chamber quantification☆European Heart Journal - Cardiovascular Imaging, 2006
- Single–suture line placement of a pericardial stentless valveThe Journal of Thoracic and Cardiovascular Surgery, 2005