Abstract
Capacitive micromachined ultrasonic transducers (CMUTs) have shown great potential to make a strong impact on ultrasound imaging technologies over the years. However, a critical weakness of CMUTs is their relatively low output pressure at transmission. To address this, we present a novel dual-element ultrasonic transducer that incorporates two transducer technologies by using a circular commercial piezoelectric element for ultrasound transmission and an annular CMUT element for reception. The hybrid transducer combines the broad bandwidth and high receive sensitivity of the CMUT and the high output pressure of the piezoelectric transducer to improve the overall sensitivity and axial resolution. The annular CMUT is designed, fabricated, and concentrically aligned with the piezoelectric probe via a custom housing. A through-wafer etching scheme based on deep reactive-ion etching is developed to create the CMUT annulus. Immersion characterization reveals that the CMUT has a center frequency of 2.63 MHz, a -6-dB fractional bandwidth of 131%, and a receive sensitivity of 22.57 mV/kPa, compared with 2.34 MHz, 42%, and 4.28 mV/kPa of the piezoelectric probe. The round-trip measurements show that the hybrid dual-element transducer improves the axial resolution by 25.58% and the signal-to-noise ratio by 8.55 dB over the commercial piezoelectric probe. [2017-0305]Keywords
Funding Information
- Natural Sciences and Engineering Research Council of Canada
- Canada Research Chairs Program
- CMC Microsystems
- China Scholarship Council
This publication has 31 references indexed in Scilit:
- A row–column addressed micromachined ultrasonic transducer array for surface scanning applicationsUltrasonics, 2014
- Capacitive micromachined ultrasonic transducers for medical imaging and therapyJournal of Micromechanics and Microengineering, 2011
- A 2D fast near-field method for calculating near-field pressures generated by apodized rectangular pistonsThe Journal of the Acoustical Society of America, 2008
- An efficient grid sectoring method for calculations of the near-field pressure generated by a circular pistonThe Journal of the Acoustical Society of America, 2004
- Fabricating capacitive micromachined ultrasonic transducers with wafer-bonding technologyJournal of Microelectromechanical Systems, 2003
- Capacitive Micromachined Ultrasonic Transducers: Theory and TechnologyJournal of Aerospace Engineering, 2003
- Characterization of a Time Multiplexed Inductively Coupled Plasma EtcherJournal of the Electrochemical Society, 1999
- Biomedical ultrasound beam formingUltrasound in Medicine & Biology, 1994
- The Evolution of Silicon Wafer Cleaning TechnologyJournal of the Electrochemical Society, 1990
- Radiation field calculations of pulsed ultrasonic transducersUltrasonics, 1980