The effect of a low radiation CT protocol on accuracy of CT guided implant migration measurement: A cadaver study
- 25 November 2015
- journal article
- research article
- Published by Wiley in Journal of Orthopaedic Research
- Vol. 34 (4) , 725-728
Abstract
The current study compared the impact of low radiation CT protocols on the accuracy, repeatability, and inter- and intra-observer variability of implant migration studies in total hip arthroplasty. Two total hip replacements were performed in two human cadavers and six tantalum beads were inserted into the femur similar to radiostereometric analysis. Six different 28 mm heads (−3 mm, 0 mm, 2.5 mm, 5.0 mm, 7.5 mm, and 10 mm) were added to simulate five reproducible translations (maximum total point migration) of the center of the head. Three CT scans with varying levels of radiation were performed for each head position. The effective dose (mSv) was 3.8 mSv for Protocol A (standard protocol), 0.7 mSv for Protocol B and 1.6 mSv for Protocol C. Implant migration was measured in a 3-D analysis software (Geomagic Studio 7). The accuracy was 0.16 mm for CT Protocol A, 0.13 mm for Protocol B and 0.14 mm for Protocol C; The repeatability was 0.22 mm for CT Protocol A, 0.18 mm for Protocol B and 0.20 mm for Protocol C; ICC for inter observer reliability was 0.89, intra observer reliability was 0.95. The difference in accuracy between standard protocol A and the two low radiation protocols (B, C) was less than 0.05 mm. The accuracy, inter- and intra-observer reliability of all three CT protocols is comparable to radiostereometric analysis. Reducing the CT radiation exposure to numbers similar to an AP Pelvis radiograph (0.7 mSv protocol B) does not affect the accuracy of implant migration measurements. Keywords: Radiation dose, CT, migration analysis, cadaver. © 2015 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 34:725–728, 2016.Keywords
This publication has 17 references indexed in Scilit:
- Imaging of Hip ArthroplastySeminars in Musculoskeletal Radiology, 2013
- Radiation Dose Modulation Techniques in the Multidetector CT Era: From Basics to PracticeRadioGraphics, 2008
- A new technique for diagnosis of acetabular cup loosening using computed tomography: Preliminary experience in 10 patientsActa Orthopaedica, 2008
- The History and Future of Radiostereometric AnalysisClinical Orthopaedics and Related Research, 2006
- The premature failure of the Charnley Elite-Plus stemThe Journal of Bone and Joint Surgery. British volume, 2006
- Use of volumetric computerized tomography as a primary outcome measure to evaluate drug efficacy in the prevention of peri‐prosthetic osteolysis: A 1‐year clinical pilot of etanercept vs. placeboJournal of Orthopaedic Research, 2003
- Acetabular component migration in total HIP arthroplasty using CT and a semiautomated program for volume mergingActa Radiologica, 2002
- acetabular component migration in total hip arthroplasty using CT and a semiautomated program for volume mergingActa Radiologica, 2002
- Accurate accuracy assessment: Review of basic principles (Acta Orthop Scand 1999; 70 (4): 319-321)Acta Orthopaedica, 2000
- Accuracy of migration analysis in hip arthroplasty Digitized and conventional radiography, compared to radiostereometry in 51 patientsActa Orthopaedica, 1995