A new design for an electron capture detector cell is described. The cell is compatible with the requirements of high resolution capillary columns and is shown to be useful in applications that require high analysis speed. A unique method of sample introduction reduces the problems of sample loss by adsorption on the surfaces of the cell. Previously reported problems of sample loss by adsorption on active surfaces within the cell when using hydrogen carrier gas are shown to have been eliminated. Examples are shown demonstrating the increased speed of analysis that can be obtained when using hydrogen carrier gas and 100 micron diameter columns.Keywords
This publication has 9 references indexed in Scilit:
- Spacial distribution of ions and electrons within 63Ni ionization cellsJournal of Chromatography A, 1982
- Increased speed of analysis in isothermal and temperature-programmed capillary gas chromatography by reduction of the column inner diameterJournal of Chromatography A, 1982
- Non-linear responses of the electron-capture detector to alkyl monochloridesJournal of Chromatography A, 1980
- Analysis errors following hydrogen cleaning of an electron capture detectorJournal of Chromatography A, 1980
- Improved model of the pulsed electron capture detectorAnalytical Chemistry, 1980
- Measurement of ions within a pulsed electron capture detector by mass spectrometryAnalytical Chemistry, 1979
- Electron-capture detectorJournal of Chromatography A, 1978
- Influence of instrumental contributions on the apparent column efficiency in high speed gas chromatographyAnalytical Chemistry, 1978
- Study in High Speed Gas Chromatography I. Injections of Narrow Sample PlugsJournal of Chromatographic Science, 1977