Optimizing rod window width in positron emission tomography
Authors:
- W.M. DigbyWDW.M. Digby
- W.K. LukWLW.K. Luk
- M.E. CaseyMCM.E. Casey
- L.G. ByarsLBL.G. Byars
- 1 June 1995
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 14 (2) , 266-270
Abstract
A technique determines the optimal window width for orbiting rod transmission studies in positron emission tomography (PET). Windowing reduces noise in orbiting rod transmission studies. Lines-of-response (LOR) which intersect the rods generate primarily true coincidence events. LOR which pass far from the rods generate random and scatter events. Since the angular position of the orbiting rods is known in real-time, LOR which produce mostly noise are gated off. When optimally determined, the rod window width maximizes the noise equivalent counts (NEC) collected in the transmission study. Transaxial rod projection profiles of trues, randoms, and scatter produce NEC versus window width plots. For the ECAT EXACT line of PET systems and a 20-cm water cylinder, optimal is five LOR wide.Keywords
This publication has 9 references indexed in Scilit:
- Performance Evaluation of the Positron Scanner ECAT EXACTJournal of Computer Assisted Tomography, 1992
- Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents countsIEEE Transactions on Nuclear Science, 1990
- Simultaneous transmission and emission scans in positron emission tomographyIEEE Transactions on Nuclear Science, 1989
- Post-injection transmission attenuation measurements for PETIEEE Transactions on Nuclear Science, 1988
- Orbiting transmission source for positron tomographyIEEE Transactions on Nuclear Science, 1988
- A Technique To Reject Scattered Radiation In PET Transmission Scans1986
- Correction for Scattered Radiation in a Ring Detector Positron Camera by Integral Transformation of the ProjectionsJournal of Computer Assisted Tomography, 1983
- Sampling Properties of Stationary and Half-Rotation Rings in Positron Emission TomographyJournal of Computer Assisted Tomography, 1981
- Data Acquisition, Reconstruction and Display for the Donner 280-Crystal Positron TomographIEEE Transactions on Nuclear Science, 1980