MR Imaging of Extracapsular Silicone from Breast Implants

Authors:

Abstract
OBJECTIVE. We sought to identify pitfalls in recognition of extracapsular silicone on MR imaging.MATERIALS AND METHODS. Three experienced observers reviewed MR images from 359 women with current (n = 320), prior (n = 15), or both current and prior (n = 24) silicone gel implants. Axial and sagittal fast spin-echo T2-weighted images with water suppression, axial inversion-recovery T2-weighted images with water suppression, and axial T2-weighted images with silicone suppression were obtained in a dedicated phased array breast coil on a 1.5-T magnet. Images were reviewed again when only one observer saw extracapsular silicone, and reasons for disagreement were recorded.RESULTS. Rupture was identified in 265 women (77%) with current silicone implants and 378 (55%) of 687 implants. Observers agreed in describing extracapsular silicone in 85 (12%) of 687 breasts with current silicone gel implants, of which 81 (95%) showed definite evidence of rupture on MR imaging. One observer reported extracapsular silicone in...

This publication has 25 references indexed in Scilit: