The closing velocity of mechanical heart valve leaflets
Authors:
- Tien-Hon ChiangTCTien-Hon Chiang
- Rodolfo C. QuijanoRQRodolfo C. Quijano
- Ned H.C. HwangNHNed H.C. Hwang
- 1 November 1994
- journal article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 16 (6) , 458-464
Abstract
The closing motion of the occluder leaflets in bileaflet type mechanical heart valves (MHV) was monitored with a laser sweeping technique. The angular displacements of the leaflets were registered with precision of 0.2 microsecond steps. Experimental measurements were made using five 29 mm Edwards-Duromedics including three original specification (EDOS) and two modified specification (EDMS), and two 29 mm St Jude Medical MHVs. The testing valve was installed in the mitral position of a physiologic pulsatile mock circulatory flow loop using water-glycerine solution as the testing fluid. Each valve was tested by: (1) direct mounting the valve on metal washers, and (2) mounting the valve with its sewing ring. Experiments were carried out at pulse rates of 70, 90, and 120 beats min-1, with the corresponding cardiac output of 5, 6, and 7.5 litres min-1, and maximum left ventricular pressure gradients (dp/dt) of 1,800, 3,000 and 5,600 mm Hg s-1, respectively. The maximum leaflet closing velocity of each of the tested valve types are presented. The difference in leaflet closing movements between the direct rigid mounting and the sewing ring mounting are discussed. The details of the laser sweeping technique are presented.Keywords
This publication has 3 references indexed in Scilit:
- Static and Dynamic Stresses during Valve Closure of a Bileaflet Mechanical Heart Valve ProsthesisThe International Journal of Artificial Organs, 1991
- Laser assessment of leaflet closing motion in prosthetic heart valvesJournal of Biomedical Engineering, 1990
- Leaflet fracture in Edwards-Duromedics bileaflet valvesThe Journal of Thoracic and Cardiovascular Surgery, 1989