Evaluation of the agreement of tidal breathing parameters measured simultaneously using pneumotachography and structured light plethysmography
Open Access
- 13 February 2017
- journal article
- research article
- Published by Wiley in Physiological Reports
- Vol. 5 (3) , e13124
Abstract
Structured light plethysmography (SLP) is a noncontact, noninvasive, respiratory measurement technique, which uses a structured pattern of light and two cameras to track displacement of the thoraco–abdominal wall during tidal breathing. The primary objective of this study was to examine agreement between tidal breathing parameters measured simultaneously for 45 sec using pneumotachography and SLP in a group of 20 participants with a range of respiratory patterns (“primary cohort”). To examine repeatability of the agreement, an additional 21 healthy subjects (“repeatability cohort”) were measured twice during resting breathing and once during increased respiratory rate (RR). Breath‐by‐breath and averaged RR, inspiratory time (tI), expiratory time (tE), total breath time (tTot), tI/tE, tI/tTot, and IE50 (inspiratory to expiratory flow measured at 50% of tidal volume) were calculated. Bland–Altman plots were used to assess the agreement. In the primary cohort, breath‐by‐breath agreement for RR was ±1.44 breaths per minute (brpm). tI, tE, and tTot agreed to ±0.22, ±0.29, and ±0.32 sec, respectively, and tI/tE, tI/tTot, and IE50/IE50SLP to ±0.16, ±0.05, and ±0.55, respectively. When averaged, agreement for RR was ±0.19 brpm. tI, tE, and tTot were within ±0.16, ±0.16, and ±0.07 sec, respectively, and tI/tE, tI/tTot, and IE50 were within ±0.09, ±0.03, and ±0.25, respectively. A comparison of resting breathing demonstrated that breath‐by‐breath and averaged agreements for all seven parameters were repeatable (P > 0.05). With increased RR, agreement improved for tI, tE, and tTot (P ≤ 0.01), did not differ for tI/tE, tI/tTot, and IE50 (P > 0.05) and reduced for breath‐by‐breath (P < 0.05) but not averaged RR (P > 0.05).Keywords
Funding Information
- PneumaCare Ltd (Cambridgeshire, UK)
This publication has 31 references indexed in Scilit:
- Tidal breathing techniques2010
- Why Bland–Altman plots should use X, not (Y+X)/2 when X is a reference methodStatistics in Medicine, 2007
- Negative expiratory pressure: A new tool for evaluating lung function in children?Pediatric Pulmonology, 2003
- Tidal breath analysis for infant pulmonary function testingEuropean Respiratory Journal, 2000
- Tidal breathing analysis and response to salbutamol in awake young children with and without asthmaEuropean Respiratory Journal, 1994
- Measurement of breath amplitudes: Comparison of three noninvasive respiratory monitors to integrated pneumotachographPediatric Pulmonology, 1993
- Respiratory rate predicts cardiopulmonary arrest for internal medicine inpatientsJournal of General Internal Medicine, 1993
- Validation of respiratory inductance plethysmography (“Respitrace”®) for the measurement of tidal breathing parameters in newbornsPediatric Pulmonology, 1992
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Tidal expiratory flow patterns in airflow obstruction.Thorax, 1981