Is there still a place for the cemented titanium femoral stem?
Authors:
- Geir HallanGHGeir Hallan
- Birgitte EspehaugBEBirgitte Espehaug
- Ove FurnesOFOve Furnes
- Helge WangenHWHelge Wangen
- Paul J HølPHPaul J Høl
- Peter EllisonPEPeter Ellison
- Leif I HavelinLHLeif I Havelin
Open Access
- 29 December 2011
- journal article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 83 (1) , 1-6
Abstract
Despite the fact that there have been some reports on poor performance, titanium femoral stems intended for cemented fixation are still used at some centers in Europe. In this population-based registry study, we examined the results of the most frequently used cemented titanium stem in Norway. 11,876 cases implanted with the cemented Titan stem were identified for the period 1987-2008. Hybrid arthroplasties were excluded, leaving 10,108 cases for this study. Stem survival and the influence of age, sex, stem offset and size, and femoral head size were evaluated using Cox regression analyses. Questionnaires were sent to the hospitals to determine the surgical technique used. Male sex, high stem offset, and small stem size were found to be risk factors for stem revision, (adjusted RR = 2.5 (1.9-3.4), 3.3 (2.3-4.8), and 2.2 (1.4-3.5), respectively). Patients operated in the period 2001-2008 had an adjusted relative risk (RR) of 4.7 (95% CI: 3.0-7.4) for stem revision due to aseptic stem loosening compared to the period 1996-2000. Changes in broaching technique and cementing technique coincided with deterioration of the results in some hospitals. The increased use of small stem sizes and high-offset stems could only explain the deterioration of results to a certain degree since the year 2000. The influence of discrete changes in surgical technique over time could not be fully evaluated in this registry study. We suggest that this cemented titanium stem should be abandoned. The results of similar implants should be carefully evaluated.Keywords
This publication has 22 references indexed in Scilit:
- Design-related risk factors for revision of primary cemented stemsActa Orthopaedica, 2010
- 18 years of results with cemented primary hip prostheses in the Norwegian Arthroplasty RegisterActa Orthopaedica, 2009
- Orthopaedic bone cementThe Journal of Bone and Joint Surgery. British volume, 2008
- In vitro comparison of Refobacin-Palacos R with Refobacin Bone Cement and Palacos R + GActa Orthopaedica, 2007
- Registration completeness in the Norwegian Arthroplasty RegisterActa Orthopaedica, 2006
- Dependency issues in survival analyses of 55 782 primary hip replacements from 47 355 patientsStatistics in Medicine, 2004
- The dynamic volume changes of polymerising polymethyl methacrylate bone cementActa Orthopaedica, 2002
- The dynamic volume changes of polymerising polymethyl methacrylate bone cementActa Orthopaedica, 2002
- The type of cement and failure of total hip replacementsThe Journal of Bone and Joint Surgery. British volume, 2002
- The Norwegian Arthroplasty Register: 11 years and 73,000 arthroplastiesActa Orthopaedica, 2000