X-ray dose delivered during a longitudinal micro-CT study has no adverse effect on cardiac and pulmonary tissue in C57BL/6 mice
Authors:
- 1 May 2013
- journal article
- Published by SAGE Publications in Acta Radiologica
- Vol. 54 (4) , 435-441
Abstract
Background Micro-computed tomography (micro-CT) offers numerous advantages for small animal imaging, including the ability to monitor the same animals throughout a longitudinal study. However, concerns are often raised regarding the effects of X-ray dose accumulated over the course of the experiment. Purpose To scan C57BL/6 mice multiple times per week for 6 weeks, in order to determine the effect of the cumulative dose on pulmonary and cardiac tissue at the end of the study. Material and Methods C57BL/6 male mice were split into two groups (irradiated group = 10, control group = 10). The irradiated group was scanned (80 kVp/50mA) three times weekly for 6 weeks, resulting in a weekly dose of 0.84 Gy, and a total study dose of 5.04 Gy. The control group was scanned on the final week. Scans from week 6 were reconstructed and the lungs and heart were analyzed. Results Overall, there was no significant difference in lung volume or lung density or in left ventricular volume or ejection fraction between the control group and the irradiated group. Histological samples taken from excised lung and myocardial tissue also showed no evidence of inflammation or fibrosis in the irradiated group. Conclusion This study demonstrated that a 5 Gy X-ray dose accumulated over 6 weeks during a longitudinal micro-CT study had no significant effects on the pulmonary and myocardial tissue of C57BL/6 mice. As a result, the many advantages of micro-CT imaging, including rapid acquisition of high-resolution, isotropic images in free-breathing mice, can be taken advantage of in longitudinal studies without concern for negative dose-related effects.Keywords
This publication has 37 references indexed in Scilit:
- Evaluation of eXIA 160XL cardiac‐related enhancement in C57BL/6 and BALB/c mice using micro‐CTContrast Media & Molecular Imaging, 2012
- Current Status of Developments and Applications of Micro-CTAnnual Review of Biomedical Engineering, 2011
- Longitudinal Follow-up of Cardiac Structure and Functional Changes in an Infarct Mouse Model Using Retrospectively Gated Micro-Computed TomographyInvestigative Radiology, 2008
- The mouse ascending: perspectives for human-disease modelsNature Cell Biology, 2007
- Cardiac Micro–Computed Tomography for Morphological and Functional Phenotyping of Muscle LIM Protein Null MiceMolecular Imaging, 2007
- Fast Retrospectively Gated Quantitative Four-Dimensional (4D) Cardiac Micro Computed Tomography Imaging of Free-Breathing MiceInvestigative Radiology, 2007
- Monte Carlo simulations of dose from microCT imaging procedures in a realistic mouse phantomMedical Physics, 2005
- 4-D Micro-CT of the Mouse HeartMolecular Imaging, 2005
- Small-Animal X-ray Dose from Micro-CTMolecular Imaging, 2004
- Initial sequencing and comparative analysis of the mouse genomeNature, 2002