The upgrade of the Thomson scattering system for measurement on the C-2/C-2U devices
Authors:
- 11 July 2016
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 87 (11) , 11D602
Abstract
The C-2/C-2U Thomson scattering system has been substantially upgraded during the latter phase of C-2/C-2U program. A Rayleigh channel has been added to each of the three polychromators of the C-2/C-2U Thomson scattering system. Onsite spectral calibration has been applied to avoid the issue of different channel responses at different spots on the photomultiplier tube surface. With the added Rayleigh channel, the absolute intensity response of the system is calibrated with Rayleigh scattering in argon gas from 0.1 to 4 Torr, where the Rayleigh scattering signal is comparable to the Thomson scattering signal at electron densities from 1 × 1013 to 4 × 1014 cm−3. A new signal processing algorithm, using a maximum likelihood method and including detailed analysis of different noise contributions within the system, has been developed to obtain electron temperature and density profiles. The system setup, spectral and intensity calibration procedure and its outcome, data analysis, and the results of electron temperature/density profile measurements will be presented.Keywords
This publication has 5 references indexed in Scilit:
- A high performance field-reversed configurationa)Physics of Plasmas, 2015
- Electron density and temperature profile diagnostics for C-2 field reversed configuration plasmasReview of Scientific Instruments, 2012
- Development of a nine spatial point, multipulse Thomson scattering diagnosticReview of Scientific Instruments, 2010
- Laser Rayleigh scatteringMeasurement Science and Technology, 2001
- The incoherent scattering of radiation from a high temperature plasmaPlasma Physics, 1972