Abstract
The commercially available REM500 neutron survey meter is a microdosimetric instrument based on the use of a low-pressure tissue-equivalent proportional counter. The ability to measure proper microdosimetric spectra, as well as the dose equivalent response, of the REM500 was evaluated with five radioisotopic neutron sources (238PuLi, 238PuF, 238PuB, 238PuBe and 252Cf) and the monoenergetic 14 MeV neutrons. The response (counts.mSv-1) and sensitivity (counts per neutron) as a function of average neutron energy were determined. Based on the measured pulse height spectra, the frequency-mean and dose-mean lineal energy values, as well as the mean quality factors, were derived for all sources. The neutron dose equivalent determined using both the dose and mean quality factor measured by the REM500 agreed well with the neutron sources' reference values of ICRP 21 maximum dose equivalent. However, for the REM500 to measure proper lineal spectra in accelerator fields, the low sensitivity and inaccurate measurements for high energy neutrons are areas that may need improvement.