Electrosurgical safety during laparoscopic surgery
Authors:
- 1 January 1995
- journal article
- research article
- Published by Taylor & Francis in Minimally Invasive Therapy
- Vol. 4 (4) , 195-201
Abstract
Dissection and haemostasis using electrosurgical current has been used safely for nearly 70 years. It is used most commonly in the monopolar form in which the circuit is completed across the patient between an active instrument and a low energy density return plate and using two current waveforms (cut and coagulation) which are manipulated to produce a variety of tissue effects. In surgical laparoscopy the monopolar arrangement may create abnormal return paths to the patient plate by capacitive and direct coupling. If these paths have a cross-sectional area sufficient to raise the power density above 7.5 watts/cm2 tissue damage may occur. These risks can be reduced by modification of surgical technique and using instrumentation designed to eliminate the abnormal return paths.Keywords
This publication has 22 references indexed in Scilit:
- Essential principles of electrosurgery in operative laparoscopyThe Journal of the American Association of Gynecologic Laparoscopists, 1994
- Unrecognized hazards of surgical electrodes passed through metal suction-irrigation devicesSurgical Endoscopy, 1994
- Bile duct injury and bile leakage in laparoscopic cholecystectomyBritish Journal of Surgery, 1993
- Diathermy in laparoscopic surgeryBritish Journal of Surgery, 1993
- Exposure, dissection, and laser versus electrosurgery in laparoscopic cholecystectomyThe American Journal of Surgery, 1993
- The laparoscopic buck stops here!The American Journal of Surgery, 1993
- Laser vs Electrosurgery in Laparoscopic CholecystectomyArchives of Surgery, 1993
- Education and engineering solutions for potential problems with laparoscopic monopolar electrosurgeryThe American Journal of Surgery, 1992
- Laser or electrocautery for laparoscopic cholecystectomy?The American Journal of Surgery, 1991
- BovieAnnals of Plastic Surgery, 1979