Electrosurgical safety during laparoscopic surgery

Authors:

Abstract
Dissection and haemostasis using electrosurgical current has been used safely for nearly 70 years. It is used most commonly in the monopolar form in which the circuit is completed across the patient between an active instrument and a low energy density return plate and using two current waveforms (cut and coagulation) which are manipulated to produce a variety of tissue effects. In surgical laparoscopy the monopolar arrangement may create abnormal return paths to the patient plate by capacitive and direct coupling. If these paths have a cross-sectional area sufficient to raise the power density above 7.5 watts/cm2 tissue damage may occur. These risks can be reduced by modification of surgical technique and using instrumentation designed to eliminate the abnormal return paths.