Pacemaker-driven Tachycardia Induced by Electrocardiograph Monitoring in the Recovery Room
Authors:
Open Access
- 1 April 2006
- journal article
- case report
- Published by SAGE Publications in Anaesthesia and Intensive Care
- Vol. 34 (2) , 266-268
Abstract
Monitoring devices are known to induce tachycardia in minute-ventilation rate-responsive pacemakers. This is because some monitoring devices measure the same parameter as do the pacemakers (change in thoracic impedance). Hence, the biological signal to the pacemaker is increased and is misconstrued as increased minute ventilation causing tachycardia which resolves when the monitoring device is removed. Whilst this could occur for all minute-volume rate responsive pacemakers, most reported interactions have been with the Telectronics META series. We present a case of an interaction between a Telectronics Tempo DR pacemaker (St. Jude Medical) and an Agilent Patient Care System (Philips). Failure to recognise the true nature and cause of such tachycardias may lead to mismanagement of the patient, including the inappropriate use of cardio-active medications.Keywords
This publication has 8 references indexed in Scilit:
- Pacemaker Misinformation in the Perioperative Period: Programming Around the ProblemAnesthesia & Analgesia, 2004
- ACC/AHA Guideline Update for Perioperative Cardiovascular Evaluation for Noncardiac Surgery—Executive SummaryAnesthesia & Analgesia, 2002
- Monitoring equipment induced tachycardia in patients with minute ventilation rate-responsive pacemakersBritish Journal of Anaesthesia, 2000
- Pacemaker-Driven Tachycardia RevisitedAnesthesia & Analgesia, 1999
- Supraventricular Tachycardia Induced by Datex Patient Monitoring SystemAnesthesia & Analgesia, 1998
- Response to “Rate-adaptive Cardiac Pacing”Anesthesiology, 1997
- Inappropriate Rate Change in Minute Ventilation Rate Responsive Pacemakers Due to Interference by Cardiac MonitorsPacing and Clinical Electrophysiology, 1997
- Rate Responsive Pacing Using a Minute Ventilation SensorPacing and Clinical Electrophysiology, 1988