Safety Design and Control Algorithm for Robotic Spinal Surgical System
Authors:
- Haiyang JinHJHaiyang Jin
- Ying HuYHYing Hu
- Feng LiFLFeng Li
- Jianwei ZhangJZJianwei Zhang
- 1 November 2011
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 23769793,p. 190-194
Abstract
This paper is based on a new Robotic Spinal Surgical System (RSSS) we have developed in last 2 years. This robotic surgical system focuses on pedicle screw implanting surgery, but can be used on some other orthopedic surgeries. Some safety design principles are presented during development of the RSSS. In this paper, two main aspects of them based on force messages are described. One is admittance control, the other is safety control during drilling process. We perform experiments to build safety control experience formula and test its availability. Our works here will support subsequent clinical practice of the RSSS.Keywords
This publication has 8 references indexed in Scilit:
- Design and kinematic analysis of A Pedicle Screws Surgical Robot2010
- Automatic identification of otological drilling faults: an intelligent recognition algorithmThe International Journal of Medical Robotics and Computer Assisted Surgery, 2010
- Design and Development of a Da Vinci Surgical System Simulator2007
- Haptic Virtual Fixtures for Robot-Assisted Manipulation2007
- MIMO State Feedback Controller for a Flexible Joint Robot with Strong Joint Coupling2007
- Development of a dexterous minimally-invasive surgical system with augmented force feedback capability2005
- Integration of UML in human factors analysis for safety of a medical robot for tele-echography2003
- Haptic control of the master hand controller for a microsurgical telerobot system2003