A CMOS Analog Correlator-Based Painless Nonenzymatic Glucose Sensor Readout Circuit
Authors:
- 16 January 2014
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Sensors Journal
- Vol. 14 (5) , 1591-1599
Abstract
An analog correlator-based readout circuit for amperometric glucose electrochemical biosensors is designed and implemented in 0.35- μm CMOS process. The readout circuit offers lower noise and wider dynamic range compared with the conventional transimpedance amplifier architecture. The operational amplifier achieves a 118.3-dB gain and 47.3 ° phase margin. The readout circuit dc response for input range of 5-30 μA gives a linear output voltage range of 0.21-1.38 V. The total output rms noise is measured to be 6.6 m Vrms. It is applied for a microneedle glucose sensor and has a mean error of 1.6% and consumes 5.1 mW of power. The total area of the proposed readout circuit is 300 μm× 167 μm.Keywords
This publication has 9 references indexed in Scilit:
- A novel microneedle-based non- enzymatic glucose sensor for painless diabetes testing application2011
- In Vivo Glucose Sensing2009
- A novel wireless glucose sensor employing direct electron transfer principle based enzyme fuel cellBiosensors and Bioelectronics, 2007
- A low noise readout circuit for integrated electrochemical biosensor arrays2006
- High aspect ratio tapered hollow metallic microneedle arrays with microfluidic interconnectorMicrosystem Technologies, 2006
- Electrochemical non-enzymatic glucose sensorsAnalytica Chimica Acta, 2006
- In-device enzyme immobilization: wafer-level fabrication of an integrated glucose sensorSensors and Actuators B: Chemical, 2004
- Nonenzymatic Glucose Detection Using Mesoporous PlatinumAnalytical Chemistry, 2003
- In vivo experiment with the electrocatalytic glucose sensor in sheepBiosensors and Bioelectronics, 1993