The mechanical test as a supplemental decision support tool for the safe removal of an Ilizarov circular external fixator
Open Access
- 7 February 2025
- journal article
- research article
- Published by Springer Nature in European Journal of Medical Research
- Vol. 30 (1) , 1-9
Abstract
Background: Timing the fixator removal is vital for a successful external fixation treatment. The purpose of this study was to determine the effectiveness of axial load–share ratio in vivo as a supplemental decision support tool for the safe removal of an Ilizarov circular external fixator. Methods: This prospective observational study consists of 83 patients undergoing tibial or femoral lengthening with Ilizarov circular external fixation in our institution, from January 2011 to October 2019. In group I (38 patients), the external fixator was removed based on the surgeon’s clinical experience and radiographs from January 2011 to June 2015. In group II (45 patients), from July 2015 to October 2019, the supplemental axial load–share (LS) ratio test was accomplished without the knowledge of the clinical results by another medical team. The test was performed by electronically measuring forces in the fixator rods and in a ground force plate. When the LS ratio < 10% was consistent with the conclusion (dense bone formation was achieved in the distraction zone) drawn from the corresponding routine radiographs by the treating surgeon, the external fixator was removed. Results: There was no statistical significance in demographic data between the two groups (P > 0.05). In group I, 4 of the 38 patients suffered refracture (the refracture rate was 10.5%) after fixator removal, and bone union was finally achieved with further intervention by intramedullary nail. In group II, 36 patients terminated the external fixation after the first mechanical test, and another 9 patients terminated the external fixation at the subsequent test. None of the 45 patients in group II suffered refracture (the refracture rate was 0%). There was statistical significance in the refracture rate between the two groups (P < 0.05). Conclusions: Adequate assessment of bone regenerate is crucial before removing an external fixator to prevent deformation or refracture. The axial load–share ratio in vivo is a practically quantitative method to supplement radiography and clinical experience for the assessment of regenerate healing, and the axial load–share ratio dropped below 10% is a safe limit for the Ilizarov circular external fixator removal.Keywords
All Related Versions
Funding Information
- the Doctoral Research Initiation Fund of Affiliated Hospital of Southwest Medical University (22046)
- the Foundation of Science and Technology plan of Sichuan (2022YFS0628)
This publication has 42 references indexed in Scilit:
- Daily subcutaneous Teriparatide injection increased bone mineral density of newly formed bone after tibia distraction osteogenesis, a randomized studyInjury, 2019
- Bone transport versus acute shortening for the management of infected tibial non-unions with bone defectsInjury, 2017
- Kallusdistraktion und Segmenttransport zur Behandlung von KnochendefektenDer Orthopäde, 2017
- Virtual stress testing of fracture stability in soldiers with severely comminuted tibial fracturesJournal of Orthopaedic Research, 2016
- In Vivo Mechanical Characterization of the Distraction Callus During Bone ConsolidationAnnals of Biomedical Engineering, 2015
- Accuracy of high-resolution in vivo micro magnetic resonance imaging for measurements of microstructural and mechanical properties of human distal tibial boneJournal of Bone and Mineral Research, 2010
- Monitoring In Vivo Load Transmission Through an External FixatorAnnals of Biomedical Engineering, 2010
- Quantitative assessment of mineralization in distraction osteogenesisSkeletal Radiology, 2008
- Transosseous application of low-intensity ultrasound for the enhancement and monitoring of fracture healing process in a sheep osteotomy modelBone, 2006
- Design and performance of a fracture monitoring transducerJournal of Biomedical Engineering, 1988