Explant pathology study of decellularized carotid artery vascular grafts
- 17 February 2004
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research Part A
- Vol. 69A (2) , 197-204
Abstract
The purpose of this study was to evaluate the morphologic findings in small-diameter freeze-dried decellularized carotid artery grafts implanted in goats as carotid artery interposition grafts for 6–7 months. Unimplanted decellularized carotid artery grafts did not contain intact cells; however, remnants of smooth muscle cells were present in the media. The extracellular matrix was well preserved. All decellularized grafts were patent at explant, without significant dimensional changes or aneurysm formation. Their luminal surfaces were lined by a thin neointima, consisting of myofibroblasts, collagen, and a discontinuous layer of endothelial cells. Histologic evidence of calcification within the explants was not observed; however, electron microscopy showed calcification of minute remnants of cell membranes. Inflammatory cells were not present in the graft wall. Host cell migration was greatest in the adventitia along the length of the graft. Migration of host cells into the media was more apparent close to the anastomoses, forming cellular nests rich in extracellular proteoglycans, whereas cell migration into areas subjacent to the lumen was minimal. Ingrowth of host blood vessels was not observed. These results demonstrate satisfactory structural and morphologic features of a decellularized carotid artery small-diameter graft implanted for up to 7 months. © 2004 Wiley Periodicals, Inc. J Biomed Mater Res 69A: 197–204, 2004Keywords
This publication has 16 references indexed in Scilit:
- Tibial bypass for limb salvage using polytetrafluoroethylene and a distal vein patchJournal of Vascular Surgery, 2001
- The radial artery versus the saphenous vein graft in contemporary CABG: a case-matched studyThe Annals of Thoracic Surgery, 2001
- A comparative study of alternative conduits for lower extremity revascularization: All-autogenous conduit versus prosthetic graftsJournal of Vascular Surgery, 2000
- The consequences of a failed femoropopliteal bypass grafting: Comparison of saphenous vein and PTFE graftsJournal of Vascular Surgery, 2000
- Intraoperative hemodynamic assessment of gastroepiploic artery and saphenous vein bypass grafts: A comparative studyThe Journal of Thoracic and Cardiovascular Surgery, 1999
- Histopathologic Findings in Synthetic and Biologic Explanted Grafts Used in Peripheral Arterial ReconstructionASAIO Journal, 1994
- Glycation, Glycoxidation, and Cross-Linking of Collagen by Glucose: Kinetics, Mechanisms, and Inhibition of Late Stages of the Maillard ReactionDiabetes, 1994
- The effects of aminoguanidine on the glycation (non-enzymic glycosylation) of lens proteinsExperimental Eye Research, 1990
- Influences of Ventricular Assist Device Pumping on Blood CoagulationASAIO Transactions, 1989
- Structural changes in glutaraldehyde-treated porcine heterografts used as substitute cardiac valvesThe American Journal of Cardiology, 1978