In-vivo dosimetry from total body irradiation patients (2000–2006): results and analysis
- 1 September 2008
- journal article
- research article
- Published by Springer Nature in Australasian Physics & Engineering Sciences in Medicine
- Vol. 31 (3) , 191-195
Abstract
In-vivo dosimetry is an important technique to ensure accuracy of delivered dose during total body irradiation (TBI). We present an analysis of semiconductor diode dosimetry, constituting seven years of dosimetry data from eighty-six patients who underwent total body irradiation. For lateral field irradiation, the mean exit dose, averaged over five anatomical sites (head, sternal notch, chest, abdomen and pelvis) and as a percentage of the planned dose was 95.7 % (SD 7.8 %). For AP/PA irradiation, the mean exit dose averaged over five anatomical sites and as a percentage of the planned dose was 95.5 % (SD 9.8 %). We propose a number of possible reasons for these differences, including patient setup variations, movement of the patient and diodes during treatment, imprecise placement of planned bolus material, inaccurate inhomogeneity corrections and modelling by the treatment planning system.Keywords
This publication has 5 references indexed in Scilit:
- In-vivo dosimetry by diode semiconductors in combination with portal films during TBI: reporting a 5-year clinical experienceRadiotherapy and Oncology, 1999
- Midplane dose determination during total body irradiation using in vivo dosimetryRadiotherapy and Oncology, 1998
- Physical aspects of total-body irradiation at the Middlesex Hospital (UCL group of hospitals), London 1988 - 1993: II.In vivoplanning and dosimetryPhysics in Medicine & Biology, 1996
- Dosimetry for total body irradiationRadiotherapy and Oncology, 1990
- Clinical and physical aspects of total body irradiation for bone marrow transplantation in NijmegenRadiotherapy and Oncology, 1990