Abstract
Among most challenging open issues in prosthetic research is the development of a robust bidirectional interface between a prosthesis and its user. Commercially available prosthetic systems are mechanically advanced, but they do not provide somatosensory feedback. Here, we present a novel non-invasive interface for multichannel electrotactile feedback, comprising a matrix of 24 pads, and we investigate the ability of able-bodied human subjects to localize the electrotactile stimulus delivered through the matrix. For this purpose, we tested conventional stimulation (same frequency for all pads) and a novel dual-parameter modulation scheme (interleaved frequency and intensity) designed to facilitate the spatial localization over the electrode. Electrotactile stimulation was also compared to mechanical stimulation of the same locations on the skin. Experimental results on eight able-bodied subjects demonstrated that the proposed interleaved coding substantially improved the spatial localization compared to same-frequency stimulation. The results also showed that same-frequency stimulation was equivalent to mechanical stimulation, whereas the performance with dual-parameter modulation was significantly better. These are encouraging outcomes for the application of a multichannel interface for the restoration of feedback in prosthetics. The high-resolution augmented interfaces might be used to explore novel scenarios for effective communication with the prosthesis user enabled by maximizing information transmission.Keywords
Funding Information
- Compagnia di San Paolo (2017.0559)
- ROBIN (8022-00243A)
- Independent Research Fund Denmark
This publication has 47 references indexed in Scilit:
- Restoring Natural Sensory Feedback in Real-Time Bidirectional Hand ProsthesesScience Translational Medicine, 2014
- Restoring the sense of touch with a prosthetic hand through a brain interfaceProceedings of the National Academy of Sciences of the United States of America, 2013
- Sensory feedback in upper limb prostheticsExpert Review of Medical Devices, 2013
- The role of feed-forward and feedback processes for closed-loop prosthesis controlJournal of NeuroEngineering and Rehabilitation, 2011
- Design and Evaluation of a Low-Cost Force Feedback System for Myoelectric Prosthetic HandsJPO Journal of Prosthetics and Orthotics, 2006
- Vibratory Adaptation of Cutaneous Mechanoreceptive AfferentsJournal of Neurophysiology, 2005
- Sensory-motor coordination during grasping and manipulative actionsCurrent Opinion in Neurobiology, 1992
- Closed-loop control in prosthetic systems: Historical perspectiveAnnals of Biomedical Engineering, 1980
- Afferent electrical nerve stimulation: Human tracking performance relevant to prosthesis sensory feedbackMedical & Biological Engineering & Computing, 1979
- Electrotactile two-point discrimination as a function of frequency, body site, laterality, and stimulation codesAnnals of Biomedical Engineering, 1977