Investigation of stochastic roughness effects for nanoscale grating characterization with a stand-alone EUV spectrometer
- 1 December 2022
- conference paper
- Published by SPIE-Intl Soc Optical Eng
- Vol. 12293, 30-158
Abstract
Fast and non-destructive non-imaging metrology of nanostructures is crucial for the development of integrated circuits and for the corresponding in-situ metrology within fabrication processes. Stochastic variations related to the gratings local period (line edge roughness, LER) and line width (line width roughness, LWR) are of special interest due to their key role for the minimal achievable structure size. Non-imaging metrology approaches taking these statistic variations into account are quite limited. For scatterometry, models predict a change of the grating’s diffraction efficiency according to a DebyeWaller factor but only in the non-zeroth diffraction orders. The authors perform simulations of nanoscale gratings that suggest an influence of LER and LWR on the reflectance (zeroth diffraction order efficiency) which motivate an extended study on LER and LWR measured by spectrally resolved EUV reflectometry here described as EUV spectrometry. The authors present reconstruction results of nanoscale gratings measured with a compact spectrometer utilizing extreme ultraviolet (EUV) radiation emitted by a discharged-produced plasma (DPP) EUV source. The use of two sequential spectrographs, one for the reference measurement of the source spectrum, the other one for the measurement of the spectrum after sample interaction, combined within the experimental setup allows to measure the broadband reflectance with 2% relative uncertainty of samples under various grazing incidence angles. The method offers a proven sub-nm reconstruction accuracy for critical grating parameters. Within the presented study, the measured samples are dedicated test samples, fabricated to exhibit well-defined LER and LWR at different grating periods and linewidths. In addition, the samples are also cross-characterized by the Physikalisch-Technische Bundesanstalt (PTB, Berlin). Experimental and simulative results are discussed to derive approaches to include LER and LWR as parameters in the physical model for reconstruction.Keywords
This publication has 9 references indexed in Scilit:
- Uncertainties in the reconstruction of nanostructures in EUV scatterometry and grazing incidence small-angle X-ray scatteringOptics Express, 2021
- Characterization of nanoscale gratings by spectroscopic reflectometry in the extreme ultraviolet with a stand-alone setupOptics Express, 2020
- Applicability of the Debye-Waller damping factor for the determination of the line-edge roughness of lamellar gratingsOptics Express, 2019
- Analytical modeling and three-dimensional finite element simulation of line edge roughness in scatterometryApplied Optics, 2012
- Effect of line roughness on the diffraction intensities in angular resolved scatterometryApplied Optics, 2010
- JCMsuite: An Adaptive FEM Solver for Precise Simulations in Nano-Optics2008
- Review Of CD Measurement And ScatterometryAIP Conference Proceedings, 2003
- X-Ray Interactions: Photoabsorption, Scattering, Transmission, and Reflection at E = 50-30,000 eV, Z = 1-92Atomic Data and Nuclear Data Tables, 1993
- Oxidation of siliconPhilosophical Magazine Part B, 1989