Safety and performance of a suprachoroidal sensor for telemetric measurement of intraocular pressure in the EYEMATE-SC trial
Open Access
- 12 November 2021
- journal article
- research article
- Published by BMJ in British Journal of Ophthalmology
- Vol. 107 (4) , 518-524
Abstract
Aim To investigate the safety and performance of a telemetric suprachoroidal intraocular pressure (IOP) sensor (EYEMATE-SC) and the accuracy of its IOP measurements in open angle glaucoma (OAG) patients undergoing simultaneous non-penetrating glaucoma surgery (NPGS). Methods Prospective, multicentre, open-label, single-arm, interventional clinical trial. Twenty-four eyes of 24 patients with OAG regularly scheduled for NPGS (canaloplasty or deep sclerectomy) were simultaneously implanted with an EYEMATE-SC sensor. Six-month follow-up on the sensor’s safety and performance as well as on the level of agreement between the EYEMATE-SC measurements and IOP measurements with Goldmann applanation tonometry (GAT). Results The eyes underwent canaloplasty (n=15) or deep sclerectomy (n=9) and achieved successful implantation of the sensor. No device migration, dislocation or serious device-related complications occurred. A total of 367 comparisons were included in the IOP agreement analysis. The overall mean difference between GAT and EYEMATE-SC measurements was 1.31 mm Hg (lower limit of agreement (LoA) 7.55 mm Hg; upper LoA –4.92 mm Hg). The maximum difference of 2.5 mm Hg ±3.96 (LoA 0.30–2.29) was reached on day 10 and continuously improved to an agreement of –0.15 mm Hg ±2.28 (LoA –1.24 to 0.89) after 6 months. Accordingly, the percentage of eyes within an IOP difference of ±5 mm Hg improved from 78% (day 3) to 100% (6 months). Conclusions After 6 months, the EYEMATE-SC sensor was safe and well tolerated, and allowed continual IOP monitoring. Trial registration number NCT03756662.Keywords
Funding Information
- This study received support from Implandata Ophthalmic Products GmbH, Hannover, Germany, which produced and provided the medical devices under investigation. (n/a)
This publication has 33 references indexed in Scilit:
- The Relationship between Glaucoma Medication Adherence, Eye Drop Technique, and Visual Field Defect SeverityOphthalmology, 2011
- 24-Hour IOP Telemetry in the Nonhuman Primate: Implant System Performance and Initial Characterization of IOP at Multiple TimescalesInvestigative Ophthalmology & Visual Science, 2011
- Continuous intraocular pressure monitoring with a wireless ocular telemetry sensor: initial clinical experience in patients with open angle glaucomaBritish Journal of Ophthalmology, 2011
- Kanaloplastie und transiente Vorderkammerblutung: ein prognostischer Faktor?Klinische Monatsblätter für Augenheilkunde, 2010
- Canaloplasty for primary open-angle glaucoma: long-term outcomeBritish Journal of Ophthalmology, 2010
- Diurnal Intraocular Pressure Patterns are Not Repeatable in the Short Term in Healthy IndividualsOphthalmology, 2010
- Autologous Transplantation of Retinal Pigment Epithelium–Bruch’s Membrane Complex for Hemorrhagic Age-Related Macular DegenerationInvestigative Ophthalmology & Visual Science, 2009
- Continuous Monitoring of Intraocular PressureJournal of Glaucoma, 2009
- Dynamic Contour Tonometry in Keratoconus and Postkeratoplasty EyesCornea, 2008
- Comparison of Goldmann applanation tonometry and dynamic contour tonometry in healthy and glaucomatous eyesEye, 2008