Influence of Anodal Electrode Position on Transvenous Defibrillation Efficacy in Humans: A Prospective Randomized Comparison
Authors:
- 1 September 1997
- journal article
- clinical trial
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 20 (9) , 2193-2199
Abstract
Nonthoracotomy lead systems for implantable cardioverter defibrillators (ICDs) have reduced operative mortality and morbidity as compared to epicardial lead systems but are usually associated with higher defibrillation thresholds (DFTs). The purpose of this prospective randomized trial was to investigate if the second defibrillation electrode in the left subclavian vein can increase defibrillation efficacy and decrease DFT as compared to the superior vena cava (SVC) position in nonthoracotomy lead systems for ICDs. Seventeen patients (mean age; 49.9 ± 11.3 years, mean ejection fraction; 46.1%± 15.8%) were implanted with an investigational unipolar electrode (Medtronic 13001) used as the defibrillation anode. DFT testing was started in the SVC (n = 10, group A) or the left subclavian vein (n = 7, group B), and repeated in the alternative position starting at the DFT of the initial position. Fifteen patients were eligible for analysis (group A: n = 9, group B: n = 6). With the electrode in the SVC, ventricular fibrillation could be successfully terminated in 9 out of 15 patients (60%). In the left subclavian vein the success rate was 100% (P < 0.01). Mean DFT in the SVC was 13.0 ± 5.2 J and in the left subclavian vein 10.2 ± 4.9 J. DFTs in the left subclavian vein were either lower (group A: n = 5/9, group B: n = 5/6) or equal to the results in the SVC position (P < 0.001). Thus, the left subclavian vein appears to be a superior alternative for positioning of the defibrillation anode as compared to the SVC for nonthoracotomy lead systems using two separate leads.Keywords
This publication has 12 references indexed in Scilit:
- Bipolar Transvenous Defibrillation: Efficacy of Two Different Positions of the AnodePacing and Clinical Electrophysiology, 1995
- The Left Subclavian Vein as an Alternative Site for Implantation of the Second Defibrillation LeadPacing and Clinical Electrophysiology, 1995
- Clinical Experience with a New Cardioverter Defibrillator Capable of Biphasic Waveform Pulse and Enhanced Data Storage: Results of a Prospective Multicenter StudyPacing and Clinical Electrophysiology, 1994
- Comparison of Simultaneous Versus Sequential Defibrillation Pulsing Techniques Using a Nonthoracotomy SystemPacing and Clinical Electrophysiology, 1994
- Early Postoperative Increase in Defibrillation Threshold with Nonthoracotomy System in HumansPacing and Clinical Electrophysiology, 1994
- Prospective, randomized comparison in humans of a unipolar defibrillation system with that using an additional superior vena cava electrode.Circulation, 1994
- Why Do Some Patients Have High Defibrillation Thresholds at Defibrillator Implantation?Pacing and Clinical Electrophysiology, 1994
- Defibrillation Thresholds and Perioperative Mortality Associated with Endocardial and Epicardial Defihrillation Lead SystemsPacing and Clinical Electrophysiology, 1993
- Three-dimensional potential gradient fields generated by intracardiac catheter and cutaneous patch electrodes.Circulation, 1992
- Current Concepts for Selecting the Location, Size and Shape of Defibrillation ElectrodesPacing and Clinical Electrophysiology, 1991