Comparison between contrast-guided and pressure-guided ablation using the novel pressure visualization tool for cryoballoon pulmonary vein isolation
Open Access
- 8 July 2025
- journal article
- research article
- Published by Springer Nature in Heart and Vessels
- Vol. 40 (12) , 1116-1124
Abstract
During cryoballon pulmonary vein isolation (PVI) complete occlusion of the pulmonary vein ostia during the freeze cycles is mandatory. Typically, PV occlusion is assessed by contrast injection under fluoroscopy. Using an update for the Cryo Console it is possible to directly visualize occlusion pressure as an indicator of complete PV occlusion during cryoballoon procedures. In this study, we compared PV pressure monitoring during cryoballoon PVI to a conventional approach regarding procedural outcomes. We retrospectively analysed the procedural data of 50 patients (25 patients with pressure-guided PVI and 25 patients with contrast-guided PVI) treated with cryoballoon PVI in our centre. Complete PV occlusion in the pressure-guided group was defined as an abrupt change in the pressure waveform with a loss of the a-wave after advancing the cryoballoon to the PV ostium. We observed comparable results regarding procedural time, left atrial dwell time or fluoroscopy time when comparing the pressure guided to our conventional approach. Moreover, there were no differences regarding acute procedural effectivity or freeze cycle characteristics. As expected, a significant reduction of contrast use was achieved in the pressure measurement group (10.4 vs. 25.5 ml, p < 0.0001). Monitoring complete PV occlusion by visualizing the occlusion pressure is feasible. Acute procedural outcome was comparable to our standard approach using contrast injection to verify complete PV occlusion. Most importantly, a significant reduction in contrast use could be achieved which has to be confirmed in larger patient cohorts.Keywords
Funding Information
- Herzzentrum Göttingen
This publication has 26 references indexed in Scilit:
- Cryoballoon ablation for atrial fibrillation without the use of a contrast medium: a combination of the intracardiac echocardiography and pressure wave monitoring guided approachHeart and Vessels, 2021
- Cryoablation or Drug Therapy for Initial Treatment of Atrial FibrillationNew England Journal of Medicine, 2021
- Cryoballoon Ablation as Initial Therapy for Atrial FibrillationNew England Journal of Medicine, 2021
- Initial experience and procedural efficacy of pulmonary vein isolation using the fourth-generation cryoballoon – a step forward?Acta Cardiologica, 2019
- Outcomes following cryoballoon ablation for atrial fibrillation guided by pressure waveform monitoring without the routine use of pulmonary venographyJournal of Interventional Cardiac Electrophysiology, 2017
- Pressure-Guided cryoablation of pulmonary veins in atrial fibrillation: A fast and effective strategyIndian Heart Journal, 2017
- Cryoablation of atrial fibrillationJournal of Interventional Cardiac Electrophysiology, 2011
- Cryoballoon temperature predicts acute pulmonary vein isolationHeart Rhythm, 2011