Background— New insight into the 3D dynamic behavior of the mitral valve has prompted a reevaluation of annuloplasty ring designs. Force balance analysis indicates correlation between annulus forces and stresses in leaflets and chords. Improving this stress distribution can intuitively enhance the durability of mitral valve repair. We tested the hypothesis that saddle-shaped annuloplasty rings have superior uniform systolic force distribution compared with a nonuniform force distribution in flat annuloplasty rings. Methods and Results— Sixteen 80-kg pigs had a flat (n=8) or saddle-shaped (n=8) mitral annuloplasty ring implanted. Mitral annulus 3D dynamic geometry was obtained with sonomicrometry before ring insertion. Strain gauges mounted on dedicated D-shaped rigid flat and saddle-shaped annuloplasty rings provided the intraoperative force distribution perpendicular to the annular plane. Average systolic annular height to commissural width ratio before ring implantation was 14.0%±1.6%. After flat and saddle shaped ring implantation, the annulus was fixed in the diastolic (9.0%±1.0%) and systolic (14.3%±1.3%) configuration, respectively ( P P Conclusions— Saddle-shaped annuloplasty rings provide superior uniform annular force distribution compared to flat rings and appear to represent a configuration that minimizes out-of-plane forces that could potentially be transmitted to leaflets and chords. This may have important implications for annuloplasty ring selections.Keywords
This publication has 14 references indexed in Scilit:
- Importance of preservation of chordal apparatus in mitral valve replacementExpert Review of Cardiovascular Therapy, 2006
- Strut chordal-sparing mitral valve replacement preserves long-term left ventricular shape and function in pigsThe Journal of Thoracic and Cardiovascular Surgery, 2005
- Epidermal Growth Factor Receptor Blockade Mediates Smooth Muscle Cell Apoptosis and Improves Survival in Rats With Pulmonary HypertensionCirculation, 2005
- Imbalanced chordal force distribution causes acute ischemic mitral regurgitation: Mechanistic insights from chordae tendineae force measurements in pigsThe Journal of Thoracic and Cardiovascular Surgery, 2005
- Static analysis of annuloplasty rings sutured on an annulus model of the mitral valve: comparison between the Duran ring and the Carpentier Classic ringJournal of Artificial Organs, 2004
- Aorto-mitral annular dynamicsThe Annals of Thoracic Surgery, 2003
- Effects of a Saddle Shaped Annulus on Mitral Valve Function and Chordal Force Distribution: An In Vitro StudyAnnals of Biomedical Engineering, 2003
- Durability of mitral valve repair for degenerative diseaseThe Journal of Thoracic and Cardiovascular Surgery, 1998
- Dynamic three-dimensional imaging of the mitral valve and left ventricle by rapid sonomicrometry array localizationThe Journal of Thoracic and Cardiovascular Surgery, 1996
- The “Physio-Ring”: an advanced concept in mitral valve annuloplastyThe Annals of Thoracic Surgery, 1995