Influence of Total Hip Design on Dislocation
Authors:
- Douglas E PadgettDPDouglas E Padgett
- Joseph LipmanJLJoseph Lipman
- Bruce RobieBRBruce Robie
- Bryan J NestorBNBryan J Nestor
- 1 June 2006
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 447, 48-52
Abstract
Dislocation following hip arthroplasty remains problematic. While the etiology of dislocation may be multifactorial, implant system design may play a role. Using a computer aided design program, virtual range of motion of several commonly implanted designs was performed with prosthetic interference representing impingement used as an endpoint. Implants with small diameter head size (22 mm) and a larger trunion geometry (type II taper) demonstrated impingement in flexion at less than 90 degrees , suggesting an increased risk for dislocation. To investigate this clinically, we performed a review of a consecutive series of 254 primary hip arthroplasties performed by a single surgeon using an implant with a type II taper (Biomet, Warsaw, IN). At a minimum 2 year followup, 12 patients with 12 hips (4.8%) had at least one episode of dislocation. Stratified by head size, the dislocation rates were 3.6% for 28 mm, 4.8% for 26 mm, and 18.8% for 22 mm bearings. These findings support the notion that computer aided design modeling of implant systems is useful in evaluating range of motion and this technique could be employed during the design of any new implant system.Keywords
This publication has 10 references indexed in Scilit:
- Impingement in Total Hip ArthroplastyThe Journal of Arthroplasty, 2005
- The Unstable Total Hip ReplacementClinical Orthopaedics and Related Research, 2004
- The influence of head and neck geometry on stability of total hip replacementA mechanical test studyActa Orthopaedica, 2004
- A mathematical formula to calculate the theoretical range of motion for total hip replacementJournal of Biomechanics, 2002
- Factors predisposing to dislocation after primary total hip arthroplastyThe Journal of Arthroplasty, 2002
- Virtual reality computer animation of the effect of component position and design on stability after total hip arthroplastyOrthopedic Clinics of North America, 2001
- The Effect of Femoral Component Head Size on Posterior Dislocation of the Artificial Hip Joint*Journal of Bone and Joint Surgery, 2000
- The Effect of the Orientation of the Acetabular and Femoral Components on the Range of Motion of the Hip at Different Head-Neck Ratios*Journal of Bone and Joint Surgery, 2000
- A Finite Element Analysis of Factors Influencing Total Hip DislocationClinical Orthopaedics and Related Research, 1998