Apparent Diffusion Coefficient Value Is Not Dependent on Magnetic Resonance Systems and Field Strength Under Fixed Imaging Parameters in Brain
Authors:
- Akio OguraAOAkio Ogura
- Takayuki TamuraTTTakayuki Tamura
- Masanori OzakiMOMasanori Ozaki
- Tsukasa Rt DoiTDTsukasa Rt Doi
- Koji FujimotoKFKoji Fujimoto
- Tosiaki PhD MiyatiTMTosiaki PhD Miyati
- Yukiko ItoYIYukiko Ito
- Fumie Rt MaedaFMFumie Rt Maeda
- Hiroyuki Rt TarewakiHTHiroyuki Rt Tarewaki
- Mitsuyuki Rt TakahashiMTMitsuyuki Rt Takahashi
- 1 January 2015
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Computer Assisted Tomography
- Vol. 39 (5) , 760-765
Abstract
Objective The aim of the study was to investigate the causes of apparent diffusion coefficient (ADC) measurement errors and to determine the optimal scanning parameters that are independent of the field strength and vendors of the magnetic resonance (MR) system. Materials and Methods Brain MR images of 10 healthy volunteers were scanned using 6 MR scanners of different field strengths and vendors in 2 different institutions. Ethical review board approvals were obtained for this study, and all volunteers gave their informed consents. Coefficient of variation (CV) of ADC values were compared for their differences in various MR scanners and in the scanned subjects. Results The CV of ADC values for 6 different scanners of 6 brains was 3.32%. The CV for repeated measurements in 1 day (10 scans per day) and in 10 days (scan per day for 10 days) for 1 subject was 1.72% and 2.96%, respectively (n = 5, P < 0.001). The CV of measurements for 10 healthy subjects was 5.22%. The measurement errors of the ADC values for 6 different MR units in 1 subject were higher than the intrascanner variance for the same subject but were lower than the intersubject variance for the same scanner. Conclusions The variance in the ADC values for different MR scanners is reasonably small if appropriate scanning parameters (repetition time, >3000 ms; echo time, minimum; and high enough signal-to-noise ratio of high-b diffusion-weighted image) are used.Keywords
This publication has 37 references indexed in Scilit:
- Value of apparent diffusion coefficient calculation in the differential diagnosis of parotid gland tumorsActa Radiologica, 2007
- Apparent diffusion coefficients of benign and malignant salivary gland tumors. Comparison to histopathological findingsJournal Of Neuroradiology, 2007
- Magnetic resonance imaging in prostate cancer: the value of apparent diffusion coefficients for identifying malignant nodulesBritish Journal of Radiology, 2007
- The Role of Diffusion-Weighted Imaging and the Apparent Diffusion Coefficient (ADC) Values for Breast TumorsKorean Journal of Radiology, 2007
- Malignant Astrocytic Tumors: Clinical Importance of Apparent Diffusion Coefficient in Prediction of Grade and PrognosisRadiology, 2006
- Diffusion‐weighted echo‐planar MR imaging and ADC mapping in the differential diagnosis of ovarian cystic masses: Usefulness of detecting keratinoid substances in mature cystic teratomasJournal of Magnetic Resonance Imaging, 2005
- The Added Value of the Apparent Diffusion Coefficient Calculation to Magnetic Resonance Imaging in the Differentiation and Grading of Malignant Brain TumorsJournal of Computer Assisted Tomography, 2004
- Study on the variations of the apparent diffusion coefficient in areas of solid tumor in high grade gliomasMagnetic Resonance Imaging, 2002
- Differentiation of clinically benign and malignant breast lesions using diffusion‐weighted imagingJournal of Magnetic Resonance Imaging, 2002
- In vivo diffusion‐weighted MRI of the breast: Potential for lesion characterizationJournal of Magnetic Resonance Imaging, 2002