Comparison of Glidescope Core, C-MAC Miller and conventional Miller laryngoscope for difficult airway management by anesthetists with limited and extensive experience in a simulated Pierre Robin sequence: A randomized crossover manikin study
Authors:
Open Access
- 22 April 2021
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 16 (4) , e0250369
Abstract
Background: Video laryngoscopy is an effective tool in the management of difficult pediatric airway. However, evidence to guide the choice of the most appropriate video laryngoscope (VL) for airway management in pediatric patients with Pierre Robin syndrome (PRS) is insufficient. Therefore, the aim of this study was to compare the efficacy of the Glidescope® Core™ with a hyperangulated blade, the C-MAC® with a nonangulated Miller blade (C-MAC® Miller) and a conventional Miller laryngoscope when used by anesthetists with limited and extensive experience in simulated Pierre Robin sequence. Methods: Forty-three anesthetists with limited experience and forty-three anesthetists with extensive experience participated in our randomized crossover manikin trial. Each performed endotracheal intubation with the Glidescope® Core™ with a hyperangulated blade, the C-MAC® with a Miller blade and the conventional Miller laryngoscope. “Time to intubate” was the primary endpoint. Secondary endpoints were “time to vocal cords”, “time to ventilate”, overall success rate, number of intubation attempts and optimization maneuvers, Cormack-Lehane score, severity of dental trauma and subjective impressions. Results: Both hyperangulated and nonangulated VLs provided superior intubation conditions. The Glidescope® Core™ enabled the best glottic view, caused the least dental trauma and significantly decreased the “time to vocal cords”. However, the failure rate of intubation was 14% with the Glidescope® Core™, 4.7% with the Miller laryngoscope and only 2.3% with the C-MAC® Miller when used by anesthetists with extensive previous experience. In addition, the “time to intubate”, the “time to ventilate” and the number of optimization maneuvers were significantly increased using the Glidescope® Core™. In the hands of anesthetists with limited previous experience, the failure rate was 11.6% with the Glidescope® Core™ and 7% with the Miller laryngoscope. Using the C-MAC® Miller, the overall success rate increased to 100%. No differences in the “time to intubate” or “time to ventilate” were observed. Conclusions: The nonangulated C-MAC® Miller facilitated correct placement of the endotracheal tube and showed the highest overall success rate. Our results therefore suggest that the C-MAC® Miller could be beneficial and may contribute to increased safety in the airway management of infants with PRS when used by anesthetists with limited and extensive experience.Keywords
This publication has 34 references indexed in Scilit:
- Anesthetic implications of infants with mandibular hypoplasia treated with mandibular distraction osteogenesisPediatric Anesthesia, 2012
- A randomized comparison of the GlideRite® Rigid Stylet to a malleable stylet for orotracheal intubation by novices using the GlideScope®Canadian Journal of Anesthesia/Journal canadien d'anesthésie, 2010
- The difficult pediatric airway – a review of new devices for indirect laryngoscopy in children younger than two years of agePediatric Anesthesia, 2010
- A comparison of the STORZ video laryngoscope and standard direct laryngoscopy for intubation in the Pediatric airway – a randomized clinical trialPediatric Anesthesia, 2009
- Management of the Difficult Infant Airway with the Storz Video Laryngoscope: A Case SeriesAnesthesia & Analgesia, 2009
- Airtraq optical laryngoscope intubation in a 5‐month‐old infant with a difficult airway because of Robin SequencePediatric Anesthesia, 2009
- Difficult airway management with the STORZ video laryngoscope in a child with Robin SequencePediatric Anesthesia, 2009
- Use of GlideScope Ranger in the management of a child with Treacher Collins syndrome in a developing world settingPediatric Anesthesia, 2009
- Airway management using the pediatric GlideScope® in a child with Goldenhar syndrome and atypical plasma cholinesterasePediatric Anesthesia, 2007
- Difficult tracheal intubation in obstetricsAnaesthesia, 1984