Ultrasonic-generated fluid velocity with Sovereign WhiteStar micropulse and continuous phacoemulsification
Authors:
- Mark E. SchaferMSMark E. Schafer
- 1 February 2006
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 32 (2) , 284-287
Abstract
To evaluate and compare ultrasonic turbulence created by conventional and micropulse ultrasound technology. Sonora Medical Systems, Longmont, Colorado, USA. A high-resolution digital ultrasound probe imaged the zone around a phacoemulsification tip. Doppler analysis allowed determination of flow. The fluid velocity was measured at 4 levels of ultrasound power at a constant flow, comparing the ultrasonic conditions of continuous energy to WhiteStar micropulses. In addition to the normal baseline irrigation and aspiration, fluid movement was detected directly below the phaco tip, produced by a nonlinear effect known as acoustic streaming. Acoustic streaming increased with increased phacoemulsification power for both conditions. At each of the 4 levels of power, fluid velocity away from the tip was less with micropulse technology than with continuous phacoemulsification. The demonstrated decrease in acoustic streaming flow away from the phaco tip with Sovereign WhiteStar micropulse technology compared to conventional ultrasound provides an objective explanation for clinical observations of increased stability of nuclear fragments at the tip and less turbulence in the anterior chamber during phacoemulsification. This methodology can be used to examine and compare fluid flow and turbulence under a variety of clinically relevant conditions.Keywords
This publication has 9 references indexed in Scilit:
- Measuring rectilinear flow within the anterior chamber in phacoemulsification proceduresJournal of Cataract & Refractive Surgery, 2004
- Risk factors for endothelial cell loss after phacoemulsification surgery by a junior residentJournal of Cataract & Refractive Surgery, 2004
- Effect of cataract surgery on the corneal endothelium*1Modern phacoemulsification compared with extracapsular cataract surgeryOphthalmology, 2004
- Acoustic cavitation in phacoemulsification: chemical effects, modes of action and cavitation indexUltrasound in Medicine & Biology, 2002
- Experimental studies on viscofluids for intraocular surgeryJournal of Cataract & Refractive Surgery, 1998
- Characteristics of acoustic streaming created and measured by pulsed Doppler ultrasoundIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1997
- A review of in vitro bioeffects of inertial ultrasonic cavitation from a mechanistic perspectiveUltrasound in Medicine & Biology, 1996
- Animal cells in turbulent fluids: Details of the physical stimulus and the biological responseBiotechnology Advances, 1993
- An experimental investigation of streaming in pulsed diagnostic ultrasound beamsUltrasound in Medicine & Biology, 1989