Calculation of beta‐ray dose distributions from ophthalmic applicators and comparison with measurements in a model eye
Authors:
- W. G. CrossWCW. G. Cross
- J. HokkanenJHJ. Hokkanen
- H. JärvinenHJH. Järvinen
- F. MourtadaFMF. Mourtada
- P. SipiläPSP. Sipilä
- C. G. SoaresCSC. G. Soares
- S. VynckierSVS. Vynckier
- 6 July 2001
- journal article
- Published by Wiley in Medical Physics
- Vol. 28 (7) , 1385-1396
Abstract
Dose distributions throughout the eye, from three types of beta‐ray ophthalmic applicators, were calculated using the EGS4, ACCEPT 3.0, and other Monte Carlo codes. The applicators were those for which doses were measured in a recent international intercomparison [Med. Phys. 28, 1373 (2001)], planar applicators ofandand a concaveapplicator. The main purpose was to compare the results of the various codes with average experimental values. For the planar applicators, calculated and measured doses on the source axis agreed within the experimental errors (<10%) to a depth of 7 mm forand 5 mm forAt greater distances the measured values are larger than those calculated. For the concaveapplicator, there was poor agreement among available calculations and only those calculated by ACCEPT 3.0 agreed with measured values. In the past, attempts have been made to derive such dose distributions simply, by integrating the appropriate point‐source dose function over the source. Here, we investigated the accuracy of this procedure for encapsulated sources, by comparing such results with values calculated by Monte Carlo. An attempt was made to allow for the effects of the silver source window but no corrections were made for scattering from the source backing. In these circumstances, at 6 mm depth, the difference in the results of the two calculations was 14%–18% for a planarapplicator and up to 30% for the concave applicator. It becomes worse at greater depths. These errors are probably caused mainly by differences between the spectrum of beta particles transmitted by the silver window and those transmitted by a thickness of water having the same attenuation properties.Keywords
This publication has 16 references indexed in Scilit:
- Dosimetry of beta‐ray ophthalmic applicators: Comparison of different measurement methodsMedical Physics, 2001
- Theoretical calculations of dose distributions for beta‐ray eye applicatorsMedical Physics, 1997
- Empirical Expressions for Beta Ray Point Source Dose DistributionsRadiation Protection Dosimetry, 1997
- ITS: the integrated TIGER series of electron/photon transport codes-Version 3.0IEEE Transactions on Nuclear Science, 1992
- Electron-photon Monte Carlo calculations: The ETRAN codeInternational Journal of Radiation Applications and Instrumentation. Part A. Applied Radiation and Isotopes, 1991
- Structure and Operation of the EGS4 Code System1988
- Presta: The parameter reduced electron-step transport algorithm for electron monte carlo transportNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1986
- Dosimetry of beta sources in radiotherapy. I. The beta point source dose functionPhysics in Medicine & Biology, 1982
- Variation of Beta Dose Attenuation in Different MediaPhysics in Medicine & Biology, 1968
- Distribution of Absorbed Energy around a Point Source of β RadiationScience, 1950