Electrical Modeling of an Isolated Surgical Aortic Valve Replacement for Microshock Risk Assessment

Authors:
  • EZEmanuele Zennaro
    Emanuele Zennaro
  • CPCarlo Mazzetti di Pietralata
    Carlo Mazzetti di Pietralata
  • FFFabio Fiamingo
    Fabio Fiamingo
  • GAGiovanni Luca Amicucci
    Giovanni Luca Amicucci

Abstract
This paper deals with the method and the concepts to obtain the circuit of the surgical layout and provides the most critical values of leakage current flowing through the heart of the patient during the phases of a surgical intervention. Different fault conditions to the operating room power supply system and to the involved electrical medical equipment are simulated. The obtained currents are compared with the values experimented by the recent investigations on electrical current inducing ventricular fibrillation (VF). The risk estimation is obtained by the product of the probability of VF when certain fault conditions occur and the probability of occurrence of the considered fault conditions. This in-depth investigation of the microshock probability provides the knowledge necessary to adopt suitable protection measures, additional to those already implemented for risk reduction.