This study investigates the use of pads with high dielectric constant (HDC) materials to alter electromagnetic field distributions in patients during magnetic resonance imaging (MRI). The study was performed with numerical simulations and phantom measurements. An initial proof-of-concept and validation was performed using a phantom at 64 MHz, showing increases of up to 10% in electromagnetic field when using distilled water as the high dielectric material. Additionally, numerical simulations with computational models of human anatomy were performed at 128 MHz. Results of these simulations using barium titanate (BaTiO3) beads showed a 61% increase of |B+1 | with a quadrature driven RF coil and a 64% increase with a dual-transmit array. The presence of the HDC material also allowed for a decrease of SAR up to twofold (e.g., peak 10 g-averaged SAR from 54 to 22 W/kg with a quadrature driven RF coil and from 27 to 22 W/kg with a dual-transmit array using CaTiO3 powder at 128 MHz). The results of this study show that the use of HDC pads at 128 MHz for MRI spine applications could result in improved magnetic fields within the region of interest, while decreasing SAR outside the region.
Funding Information
  • FDA Critical Path Initiation
  • Research Participation Program at the Center for Devices and Radiological Health administered
  • Oak Ridge Institute for Science and Education
  • U.S. Department of Energy
  • U.S. Food and Drug Administration

This publication has 24 references indexed in Scilit: