A Novel Hollow Active Auxiliary Electrode Applied for the Vertical Calculable Cross Capacitor at NIM
Authors:
- 5 February 2015
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 64 (6) , 1657-1662
Abstract
To compensate the end effect error in a vertical calculable cross capacitor with a fixed and a movable guard electrode, two different methods, a nosepiece and a modified active auxiliary electrode, are usually adopted at National Measurement Institutes across the world. However, the latter electrical approach can be more difficult when assembling the mirrors in the resonant cavity of the Fabry-Perot laser interferometer. This paper describes a novel hollow active auxiliary electrode that can satisfy the optical requirements. The model test experimental results show that it has the same or better compensation effectiveness for the end effect error in comparison with the modified active auxiliary electrode.Keywords
Funding Information
- Key Project in the National Science and Technology Pillar Program of China (2006BAF06B04)
- National Science Foundation of China (51207148, 51207149, 51377150)
This publication has 6 references indexed in Scilit:
- Model Tests of Electrical Compensation Method for the New Calculable Cross-Capacitor at NIMIEEE Transactions on Instrumentation and Measurement, 2013
- Fabrication of the main electrodes of the NMIA-BIPM calculable capacitor2010
- Design of a Calculable Cross-Capacitor2004
- Apparatus for Measuring the Cylindricity of Calculable Cross-Capacitor Electrodes2004
- Model tests to investigate the effects of geometrical imperfections on the NIST calculable capacitorIEEE Transactions on Instrumentation and Measurement, 1999
- A Calculable Standard of CapacitanceMetrologia, 1965