Proximal Component Modularity in THA—At What Cost?: An Implant Retrieval Study
Authors:
- C. KeoghCKC. Keogh
- E. SwartsESE. Swarts
- 1 July 2012
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 470 (7) , 1885-1894
Abstract
There have been reports of catastrophic fracture, cold welding, and corrosion and fretting of the modular junction. Questions/purposes We asked whether (1) the neck-stem junction showed the same degradation mechanisms, if any, as the head-neck junction, (2) the junction contributed to THA revision, (3) the alloy affected the degree of degradation, and (4) the trunion machine finish affected the degradation mechanisms. Methods We compared 57 retrievals from seven total hip modular designs, three cobalt-chromium-molybdenum and four titanium based: Bionik® (four), GMRS® (four), Margron® (22), Apex® (five), M-series® (five), ZMR® (two), and S-ROM® (15). Macroscopic inspection, microscopy, and micro-CT were conducted to determine the effects of materials and design. Results The cobalt-chromium-molybdenum components showed crevice corrosion and fretting of the neck-stem taper, whereas the titanium components had less corrosion; however, there were several cases of cold welding where disassembly could not be achieved in theater. Conclusions Even with modern taper designs and corrosion-resistant materials, corrosion, fretting, and particulate debris were observed to a greater extent in the second neck-stem junction. Titanium-based modular arthroplasty may lessen the degree of degradation, but cold welding of the components may occur. Clinical Relevance Degradation of the second junction contributed to 8 cases of metallosis and two cases of aseptic lymphocyte-laminated vascular-associated lesions contributing to revision....Keywords
This publication has 24 references indexed in Scilit:
- The Proximal Modular Neck in THA: A Bridge Too Far: AffirmsOrthopedics, 2010
- The Proximal Modular Neck in THA: A Bridge Too Far: OpposesOrthopedics, 2010
- The John Charnley Award: Metal-on-Metal Hip Resurfacing versus Large-diameter Head Metal-on-Metal Total Hip ArthroplastyClinical Orthopaedics and Related Research, 2010
- Metal release and corrosion effects of modular neck total hip arthroplastyInternational Orthopaedics, 2009
- Effects of soluble cobalt and cobalt incorporated into calcium phosphate layers on osteoclast differentiation and activationBiomaterials, 2009
- Modular Hip Stems: Determination of Disassembly Force of a Neck–Stem CouplingArtificial Organs, 2007
- The Morse Taper Junction in Modular Revision Hip Replacement – A Biomechanical and Retrieval Analysis - Die konische Steckverbindung in der modularen Revisionsendoprothetik der Hüfte – Vergleich eines explantierten Prothesenschaftes zur In-vitro-TestungBiomedical Engineering / Biomedizinische Technik, 2000
- Fretting wear in a modular neck hip prosthesisJournal of Biomedical Materials Research, 1997
- Design-related fretting wear in modular neck hip prosthesisJournal of Biomedical Materials Research, 1996
- Fretting corrosion studies of universal femoral head prostheses and cone taper spigotsBiomaterials, 1990