Performance assessment of the Medtronic‐MiniMed Continuous Glucose Monitoring System and its use for measurement of glycaemic control in Type 1 diabetic subjects
Authors:
- 24 November 2003
- journal article
- research article
- Published by Wiley in Diabetic Medicine
- Vol. 20 (12) , 1012-1015
Abstract
Aims To assess the accuracy, reliability and measurement of glycaemic control associated with the Medtronic‐MiniMed Continuous Glucose Monitoring System (CGMS) in comparison with blood glucose self‐monitoring (BGSM) in Type 1 diabetic patients.Methods Type 1 diabetic patients (n = 18) underwent glucose monitoring by the CGMS for up to 3 days, when control was also assessed by BGSM performed eight times daily.Results Ninety‐five per cent of paired non calibration samples were in the clinically acceptable zones of the Clarke error grid, and 97% in a consensus grid. The median bias was 0.1 mmol/l and relative bias 15%. The failure rate of sensors was 28% of those initially inserted. The CGMS detected significantly more hypoglycaemia and post‐prandial hyperglycaemia, but total duration of hyperglycaemia, blood glucose oscillations and day‐to‐day variability were similarly assessed by CGMS and BGSM.Conclusions The CGMS has acceptable clinical accuracy. Detection by the CGMS of more hypoglycaemia and post‐prandial hyperglycaemia than BGSM promises a valuable tool for control assessment and optimization of treatment.Keywords
This publication has 17 references indexed in Scilit:
- Reproducibility of Glucose Measurements Using the Glucose SensorDiabetes Care, 2002
- A Pilot Study of the Continuous Glucose Monitoring SystemDiabetes Care, 2001
- Limitations of Conventional Methods of Self-Monitoring of Blood GlucoseDiabetes Care, 2001
- Continuous Subcutaneous Glucose Monitoring in Children With Type 1 DiabetesPediatrics, 2001
- The MiniMed Continuous Glucose Monitoring SystemDiabetes Technology & Therapeutics, 2000
- Continuous Glucose Monitoring in Previously Unstudied Population SubgroupsDiabetes Technology & Therapeutics, 2000
- A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose.Diabetes Care, 2000
- Performance Evaluation of the MiniMed® Continuous Glucose Monitoring System During Patient Home UseDiabetes Technology & Therapeutics, 2000
- Continuous glucose monitoring used to adjust diabetes therapy improves glycosylated hemoglobin: a pilot studyDiabetes Research and Clinical Practice, 1999
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987