The effect of temperature on bougies: a photographic and manikin study*
Authors:
- 5 September 2008
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 63 (10) , 1135-1138
Abstract
Summary: We investigated the effect of temperature on single use and reuseable bougies. In a photographic study, three bougies (Portex ® Venn reuseable (R), Portex single use (S) and Breathesafe ™ single use (B)) were exposed to increasing temperatures and sequential photographs were taken of the bougies uncoiling from a preformed curve. Bougie type was associated with rate of uncoiling, type R maintaining its curve the longest but changing temperature did not affect this. In a randomised cross‐over manikin study, 16 anaesthetists attempted to pass two bougies (Portex reuseable (R) and Portex single use (S)) at three temperatures (10, 20 and 30 °C) into the trachea of a manikin. Type R was significantly associated with higher success rate of tracheal placement compared to type S. Change of temperature was significant with success rate increasing with lower temperature. The odds of success at 10 °C was six times that at 30 °C (OR (95%) 6.7 (1.7, 25.7)). We concluded that both bougies performed best at 10 °C.Keywords
This publication has 7 references indexed in Scilit:
- A comparison of multiple‐use and single‐use bougiesAnaesthesia, 2005
- Evaluation of Frova, single‐use intubation introducer, in a manikin. Comparison with Eschmann multiple‐use introducer and Portex single‐use introducer*Anaesthesia, 2004
- Difficult Airway Society guidelines for management of the unanticipated difficult intubationAnaesthesia, 2004
- Bougie‐assisted difficult airway management in a manikin – the effect of position held on placement and force exerted by the tip*Anaesthesia, 2004
- To shape or not to shape…Simulated bougie‐assisted difficult intubation in a manikin*Anaesthesia, 2003
- A comparison of simulated difficult intubation with multiple‐use and single‐use bougies in a manikin*Anaesthesia, 2002
- Difficult tracheal intubation in obstetricsAnaesthesia, 1984