Detection of admittivity anomaly on high-contrast heterogeneous backgrounds using frequency difference EIT
Authors:
- 26 May 2015
- journal article
- Published by IOP Publishing in Physiological Measurement
- Vol. 36 (6) , 1179-1192
Abstract
This paper describes a multiple background subtraction method in frequency difference electrical impedance tomography (fdEIT) to detect an admittivity anomaly from a high-contrast background conductivity distribution. The proposed method expands the use of the conventional weighted frequency difference EIT method, which has been used limitedly to detect admittivity anomalies in a roughly homogeneous background. The proposed method can be viewed as multiple weighted difference imaging in fdEIT. Although the spatial resolutions of the output images by fdEIT are very low due to the inherent ill-posedness, numerical simulations and phantom experiments of the proposed method demonstrate its feasibility to detect anomalies. It has potential application in stroke detection in a head model, which is highly heterogeneous due to the skull.Keywords
This publication has 29 references indexed in Scilit:
- Validation of weighted frequency-difference EIT using a three-dimensional hemisphere model and phantomPhysiological Measurement, 2011
- Electrical impedance tomography using a point electrode inverse scheme for complete electrode dataInverse Problems and Imaging, 2011
- An electrode addressing protocol for imaging brain function with electrical impedance tomography using a 16-channel semi-parallel systemPhysiological Measurement, 2009
- GREIT: a unified approach to 2D linear EIT reconstruction of lung imagesPhysiological Measurement, 2009
- Non-linear integral equations for the complete electrode model in inverse impedance tomographyApplicable Analysis, 2008
- Design and implementation of a high frequency electrical impedance tomography systemPhysiological Measurement, 2004
- The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHzPhysics in Medicine & Biology, 1996
- Measured and expected Cole parameters from electrical impedance tomographic spectroscopy images of the human thoraxPhysiological Measurement, 1995
- A dual-frequency applied potential tomography technique: computer simulationsClinical Physics and Physiological Measurement, 1987
- The specific resistance of biological material—A compendium of data for the biomedical engineer and physiologistMedical & Biological Engineering & Computing, 1967