Evaluation of the platelet counting by Abbott CELL-DYN®SAPPHIRE™ haematology analyser compared with flow cytometry
Authors:
- 1 April 2009
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 31 (2) , 151-160
Abstract
The Abbot Cell-Dyn Sapphire is a new generation haematology analyser. The system uses optical/fluorescence flow cytometry in combination with electronic impedance to produce a full blood count. Optical and impedance are the default methods for platelet counting while automated CD61-immunoplatelet analysis can be run as selectable test. The aim of this study was to determine the platelet count performance of the three counting methods available on the instrument and to compare the results with those provided by Becton Dickinson FACSCalibur flow cytometer used as reference method. A lipid interference experiment was also performed. Linearity, carryover and precision were good, and satisfactory agreement with reference method was found for the impedance, optical and CD61-immunoplatelet analysis, although this latter provided the closest results in comparison with flow cytometry. In the lipid interference experiment, a moderate inaccuracy of optical and immunoplatelet counts was observed starting from a very high lipid value.Keywords
This publication has 15 references indexed in Scilit:
- The platelet count accuracy of platelet concentrates obtained by using automated analysis is influenced by instrument bias and activated platelet componentsVox Sanguinis, 2004
- Quality specification in haematology: the automated blood cell countClinica Chimica Acta, 2004
- Reduction of the Platelet Review Rate Using the Two-Dimensional Platelet MethodAmerican Journal of Clinical Pathology, 2001
- Platelet Counting by the RBC/Platelet Ratio MethodAmerican Journal of Clinical Pathology, 2001
- An Interlaboratory Study of a Candidate Reference Method for Platelet CountingAmerican Journal of Clinical Pathology, 2001
- Automated CD61 immunoplatelet analysis of thrombocytopenic samplesBritish Journal of Haematology, 2001
- Improved Platelet Counting Using Two-Dimensional Laser Light ScatterAmerican Journal of Clinical Pathology, 2000
- Acute Immune Thrombocytopenia in Childhood. Are We Treating the Platelet Count?Seminars in Thrombosis and Hemostasis, 1998
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Recommended scheme for the evaluation of instruments for automatic analysis in the clinical biochemistry laboratory.Journal of Clinical Pathology, 1969