Intrinsically stable CCC for accurate measurement of small currents
- 1 June 2008
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 156-157
Abstract
We have developed and tested an intrinsically stable cryogenic current comparator (CCC) with a 1:20000 winding ratio for the accurate measurement of small electrical currents. Key feature of the CCC is the use of resistive CuNi wire and brass internal shields. First measurements of lOOpA to 300 pA currents using this CCC are performed at an uncertainty level of around 20 ppm to 50 ppm.Keywords
This publication has 4 references indexed in Scilit:
- Direct resistance comparisons from the QHR to100 M/spl Omega/ using a cryogenic current comparatorIEEE Transactions on Instrumentation and Measurement, 2005
- 1:30 000 cryogenic current comparator with optimum squid readoutIEEE Transactions on Instrumentation and Measurement, 2003
- Accurate measurement of currents generatedby single electronstransported in a one-dimensional channelIEE Proceedings - Science, Measurement and Technology, 2000
- A Capacitance Standard Based on Counting ElectronsScience, 1999