The effects of spatial sampling choices on MR temperature measurements
Authors:
Open Access
- 29 September 2010
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 65 (2) , 515-521
Abstract
The purpose of this article is to quantify the effects that spatial sampling parameters have on the accuracy of magnetic resonance temperature measurements during high intensity focused ultrasound treatments. Spatial resolution and position of the sampling grid were considered using experimental and simulated data for two different types of high intensity focused ultrasound heating trajectories (a single point and a 4‐mm circle) with maximum measured temperature and thermal dose volume as the metrics. It is demonstrated that measurement accuracy is related to the curvature of the temperature distribution, where regions with larger spatial second derivatives require higher resolution. The location of the sampling grid relative temperature distribution has a significant effect on the measured values. When imaging at 1.0 × 1.0 × 3.0 mm3 resolution, the measured values for maximum temperature and volume dosed to 240 cumulative equivalent minutes (CEM) or greater varied by 17% and 33%, respectively, for the single‐point heating case, and by 5% and 18%, respectively, for the 4‐mm circle heating case. Accurate measurement of the maximum temperature required imaging at 1.0 × 1.0 × 3.0 mm3 resolution for the single‐point heating case and 2.0 × 2.0 × 5.0 mm3 resolution for the 4‐mm circle heating case. Magn Reson Med, 2011.Keywords
This publication has 22 references indexed in Scilit:
- Extracorporeal high intensity focused ultrasound ablation in the treatment of 1038 patients with solid carcinomas in China: an overviewUltrasonics Sonochemistry, 2004
- MRI feedback temperature control for focused ultrasound surgeryPhysics in Medicine & Biology, 2002
- Tissue thermal conductivity by magnetic resonance thermometry and focused ultrasound heatingJournal of Magnetic Resonance Imaging, 2002
- Analysis of partial volume effects in diffusion‐tensor MRIMagnetic Resonance in Medicine, 2001
- Local hyperthermia with MR-guided focused ultrasound: Spiral trajectory of the focal point optimized for temperature uniformity in the target regionJournal of Magnetic Resonance Imaging, 2000
- Focused US system for MR imaging-guided tumor ablation.Radiology, 1995
- Noninvasive MRI Thermometry with the Proton Resonance Frequency (PRF) Method: In Vivo Results in Human MuscleMagnetic Resonance in Medicine, 1995
- Reduction of partial‐volume artifacts with zero‐filled interpolation in three‐dimensional MR angiographyJournal of Magnetic Resonance Imaging, 1994
- Accuracy of phase‐contrast flow measurements in the presence of partial‐volume effectsJournal of Magnetic Resonance Imaging, 1993
- Thermal dose determination in cancer therapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1984