Evaluating a calcium-aware kernel for CT CAC scoring with varying surrounding materials and heart rates: a dynamic phantom study
Open Access
- 28 May 2021
- journal article
- research article
- Published by Springer Nature in European Radiology
- Vol. 31 (12) , 9211-9220
Abstract
Objectives: The purpose of this study was twofold. First, the influence of a novel calcium-aware (Ca-aware) computed tomography (CT) reconstruction technique on coronary artery calcium (CAC) scores surrounded by a variety of tissues was assessed. Second, the performance of the Ca-aware reconstruction technique on moving CAC was evaluated with a dynamic phantom. Methods: An artificial coronary artery, containing two CAC of equal size and different densities (196 ± 3, 380 ± 2 mg hydroxyapatite cm−3), was moved in the center compartment of an anthropomorphic thorax phantom at different heart rates. The center compartment was filled with mixtures, which resembled fat, water, and soft tissue equivalent CT numbers. Raw data was acquired with a routine clinical CAC protocol, at 120 peak kilovolt (kVp). Subsequently, reduced tube voltage (100 kVp) and tin-filtration (150Sn kVp) acquisitions were performed. Raw data was reconstructed with a standard and a novel Ca-aware reconstruction technique. Agatston scores of all reconstructions were compared with the reference (120 kVp) and standard reconstruction technique, with relevant deviations defined as > 10%. Results: For all heart rates, Agatston scores for CAC submerged in fat were comparable to the reference, for the reduced-kVp acquisition with Ca-aware reconstruction kernel. For water and soft tissue, medium-density Agatston scores were again comparable to the reference for all heart rates. Low-density Agatston scores showed relevant deviations, up to 15% and 23% for water and soft tissue, respectively. Conclusion: CT CAC scoring with varying surrounding materials and heart rates is feasible at patient-specific tube voltages with the novel Ca-aware reconstruction technique. Key Points: • A dedicated calcium-aware reconstruction kernel results in similar Agatston scores for CAC surrounded by fatty materials regardless of CAC density and heart rate. • Application of a dedicated calcium-aware reconstruction kernel allows for radiation dose reduction. • Mass scores determined with CT underestimated physical mass.Keywords
This publication has 24 references indexed in Scilit:
- Coronary artery calcium: A technical argument for a new scoring methodJournal of Cardiovascular Computed Tomography, 2019
- Coronary Calcium as a Predictor of Coronary Events in Four Racial or Ethnic GroupsThe New England journal of medicine, 2008
- Coronary Calcification: Effect of Small Variation of Scan Starting Position on Agatston, Volume, and Mass ScoresRadiology, 2008
- The Influence of Heart Rate, Slice Thickness, and Calcification Density on Calcium Scores Using 64-Slice Multidetector Computed TomographyInvestigative Radiology, 2007
- Evidence for lower variability of coronary artery calcium mineral mass measurements by multi-detector computed tomography in a community-based cohort—Consequences for progression studiesEuropean Journal of Radiology, 2006
- Coronary Calcium Measurements: Effect of CT Scanner Type and Calcium Measure on Rescan Reproducibility—MESA StudyRadiology, 2005
- Prognostic Value of Cardiac Risk Factors and Coronary Artery Calcium Screening for All-Cause MortalityRadiology, 2003
- Coronary Artery Calcium: Accuracy and Reproducibility of Measurements with Multi–Detector Row CT—Assessment of Effects of Different Thresholds and Quantification MethodsRadiology, 2003
- Long-Term Prognostic Value of Coronary Calcification Detected by Electron-Beam Computed Tomography in Patients Undergoing Coronary AngiographyCirculation, 2001
- Quantification of coronary artery calcium using ultrafast computed tomographyJACC, 1990