Using dynamic ultrasound to assess Achilles tendon mechanics during running: the effect on running pattern and muscle-tendon junction tracking
Authors:
Preprint
Abstract
Achilles tendon strain can be quantified using dynamic ultrasound, but its use in running is limited. Minimal effects on running pattern and acceptable test-retest reliability of muscle-tendon junction (MTJ) tracking are prerequisites for ultrasound use during running. We aimed to assess (i) the effect of wearing an ultrasound transducer on running pattern and (ii) the test-retest reliability of MTJ tracking during running. Sixteen long-distance runners (nine injury-free, seven with Achilles tendinopathy) ran at different speeds on an instrumented treadmill with a 10-camera system tracking skin-mounted retroreflective markers, first without and then with an ultrasound transducer attached to the lower leg to track the MTJ of the gastrocnemius medialis. Spatiotemporal parameters, joint kinematics and kinetics were compared between conditions using mixed ANOVAs and paired t-tests. MTJ tracking was performed manually twice by three raters in ten participants. Variability and standard error of measurement (SEM) quantified the inter- and intra-tester test-retest reliability. The running pattern was not affected by wearing the ultrasound transducer, except for significantly less knee flexion during midstance (1.6°) and midswing (2.9°) found when wearing the transducer. Inter-rater and intra-rater SEMs for MTJ tracking to assess the tendon strain (0.43%, and 0.56%, respectively) were about four times as low as between-group differences presented in literature. The minimal effects found on the running pattern and acceptable test-retest reliability indicates that dynamic ultrasound during running can be appropriately used to study Achilles tendon mechanics and thereby help improve our understanding of Achilles tendon behavior during running, injury development and recovery.Keywords
All Related Versions
- Published version: Journal of Biomechanics, 176, 112344.
This publication has 22 references indexed in Scilit:
- Wearing an ultrasound probe during walking does not influence lower limb joint kinematics in adolescents with cerebral palsy and typically developing peersGait & Posture, 2024
- The influence of wearing an ultrasound device on gait in children with cerebral palsy and typically developing childrenGait & Posture, 2023
- Effects of soft tissue artifacts on the calculated kinematics of the knee during walking and runningJournal of Biomechanics, 2023
- The Human Body Model versus conventional gait models for kinematic gait analysis in children with cerebral palsyHuman Movement Science, 2020
- The reliability and validity of a clinical 3D freehand ultrasound systemComputer Methods and Programs in Biomedicine, 2016
- Gastrocnemius muscle–tendon interaction during walking in typically-developing adults and children, and in children with spastic cerebral palsyJournal of Biomechanics, 2016
- A real-time system for biomechanical analysis of human movement and muscle functionMedical & Biological Engineering & Computing, 2013
- The effects of a 30-min run on the mechanics of the human Achilles tendonEuropean Journal of Applied Physiology, 2011
- The reliability of three-dimensional kinematic gait measurements: A systematic reviewGait & Posture, 2009
- Two simple methods for determining gait events during treadmill and overground walking using kinematic dataGait & Posture, 2008