Improved Pixel-wise Calibration for Charge-Integrating Hybrid Pixel Detectors with Performance Validation
Authors:
Preprint
Abstract
The MÖNCH hybrid pixel detector, with a 25 \textmu m pixel pitch and fast charge-integrating readout, has demonstrated subpixel resolution capabilities for X-ray imaging and deep learning-based electron localization in electron microscopy. Fully exploiting this potential requires extensive calibration to ensure both linearity and uniformity of the pixel response, which is challenging for detectors with a large dynamic range. To overcome the limitations of conventional calibration methods, we developed an accurate and efficient correction method to achieve pixel-wise gain and nonlinearity calibration based on the backside pulsing technique. A three-dimensional lookup table was generated for all pixels across the full dynamic range, mapping the pixel response to a calibrated linear energy scale. Compared with conventional linear calibration, the proposed method yields negligible deviations between the calibrated and nominal energies for photons and electrons. The improvement in energy resolution ranges from 4% to 22% for 15-25 keV photons and from 16% to 23% for 60-200 keV electrons. Deep learning-based electron localization demonstrates a 4% improvement in spatial resolution when using the proposed calibration method. This approach further enables rapid diagnosis of the cause of bad pixels and estimation of bump-bonding yield.Keywords
All Related Versions
This publication has 15 references indexed in Scilit:
- Optimization and validation of charge transport simulation for hybrid pixel detectors incorporating the repulsion effectNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2025
- Calibration methods for charge integrating detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2021
- SciPy 1.0: fundamental algorithms for scientific computing in PythonNature Methods, 2020
- First full dynamic range calibration of the JUNGFRAU photon detectorJournal of Instrumentation, 2018
- Measurements with MÖNCH, a 25 μm pixel pitch hybrid pixel detectorJournal of Instrumentation, 2017
- New calibration circuitry and concept for AGIPDJournal of Instrumentation, 2016
- Characterization results of the JUNGFRAU full scale readout ASICJournal of Instrumentation, 2016
- Metrology and Tests beamline at SOLEIL Design and first resultsAIP Conference Proceedings, 2010
- Measurement of the mean electron-hole pair creation energy in crystalline silicon for photons in the 50–1500 eV spectral rangeApplied Physics Letters, 1996
- Relativistic hartree-slater values for K and L X-ray emission ratesAtomic Data and Nuclear Data Tables, 1974