Measurement of the electrical and mechanical responses of a force transducer against impact forces
Author:
- 1 August 2006
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 77 (8) , 085108
Abstract
A method for measuring the electrical and mechanical responses of force transducers to impact loads is proposed. The levitation mass method (LMM) is used to generate and measure the reference impact force used. In the LMM, a mass that is levitated using an aerostatic linear bearing (and hence encounters negligible friction) is made to collide with the force transducer under test, and the force acting on the mass is measured using an optical interferometer. The electrical response is evaluated by comparing the output signal of the force transducer with the inertial force of the mass as measured using the optical interferometer. Simultaneously, the mechanical response is evaluated by comparing the displacement of the sensing point of the transducer, which is measured using another optical interferometer, with the inertial force of the mass. To demonstrate the efficiency of the proposed method, the impact responses of a force transducer are accurately determined.Keywords
This publication has 15 references indexed in Scilit:
- Proposal for material viscoelasticity evaluation method under impact loadJournal of Materials Science, 2005
- Method for evaluating material viscoelasticityReview of Scientific Instruments, 2004
- Proposal for a step response evaluation method for force transducersMeasurement Science and Technology, 2003
- A method for calibrating force transducers against oscillation forceMeasurement Science and Technology, 2003
- Measurement of force acting on a moving part of a pneumatic linear bearingReview of Scientific Instruments, 2003
- Dynamic investigation of a binocular six-component force-moment sensorMeasurement Science and Technology, 2002
- From oscillation to impact: the design of a new force calibration device at PTBMeasurement, 2002
- Dynamic investigation of a three-component force-moment sensorMeasurement Science and Technology, 2002
- Measurement of impulse response of force transducersReview of Scientific Instruments, 2001
- Investigation of the comparison method for the dynamic calibration of force transducersMeasurement, 1998