Mass measurement accuracy of the Orbitrap in intact proteome analysis
- 15 May 2016
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 30 (12) , 1391-1397
Abstract
Rationale: The mass measurement accuracy (MMA) of Orbitrap mass spectrometers is 1–5 ppm according to the manufacturer's specification; yet, up to 50 ppm has been used as mass tolerance to interpret Orbitrap data in the literature. A systematic evaluation of MMA is thus necessary to find the optimal mass tolerance to be used. Methods: Reversed‐phase liquid chromatography/tandem mass spectrometry (RPLC/MS/MS) analyses of the intact E. coli proteome were carried out on a Q Exactive Orbitrap mass spectrometer coupled to a Dionex UltiMate 3000 RSLCnano system. The analysis included three technical replicates each day and was repeated for six continuous days right after a mass calibration. The obtained raw datasets were searched using ProteinGoggle 2.0 under four different mass tolerances of 5, 10, 15, and 20 ppm. Results: With both forward and random database searches and FDR ≤1% at the spectrum level, the most protein spectrum matches and protein IDs were obtained at a mass tolerance of 15 ppm. The average mass accuracy of both precursor and product ions from three representative high, medium, and low abundance proteins as well as the common proteins identified in all the 18 replicate runs was found to be 0–4 ppm; and no significant drift of measured mass accuracy was observed within the calibration period of 1 week. Conclusions: Despite the mass measurement accuracy of 1–5 ppm of the Orbitrap stated by the manufacturer, the optimal mass tolerance for protein identification was found to be 15 ppm for both the precursor and product ions. Weekly mass calibration is appropriate because no significant drift in MMA was found within the 6‐day period. Copyright © 2016 John Wiley & Sons, Ltd.Keywords
This publication has 14 references indexed in Scilit:
- Anion-Exchange Chromatography of Phosphopeptides: Weak Anion Exchange versus Strong Anion Exchange and Anion-Exchange Chromatography versus Electrostatic Repulsion–Hydrophilic Interaction ChromatographyAnalytical Chemistry, 2015
- Confident Site Localization Using a Simulated Phosphopeptide Spectral LibraryJournal of Proteome Research, 2015
- Middle-Down Analysis of Monoclonal Antibodies with Electron Transfer Dissociation Orbitrap Fourier Transform Mass SpectrometryAnalytical Chemistry, 2014
- High‐accuracy identification and bioinformatic analysis of in vivo protein phosphorylation sites in yeastProteomics, 2009
- Dynamic range of mass accuracy in LTQ orbitrap hybrid mass spectrometerJournal of the American Society for Mass Spectrometry, 2006
- Top-down Protein Sequencing and MS3 on a Hybrid Linear Quadrupole Ion Trap-Orbitrap Mass SpectrometerMolecular & Cellular Proteomics, 2006
- Performance Evaluation of a Hybrid Linear Ion Trap/Orbitrap Mass SpectrometerAnalytical Chemistry, 2006
- Parts per Million Mass Accuracy on an Orbitrap Mass Spectrometer via Lock Mass Injection into a C-trapMolecular & Cellular Proteomics, 2005
- Performance of a Linear Ion Trap-Orbitrap Hybrid for Peptide AnalysisAnalytical Chemistry, 2005
- Fourier transform ion cyclotron resonance spectroscopyChemical Physics Letters, 1974