Abstract
Even though digital breast tomosynthesis (DBT) is a well-established imaging modality, efforts to optimize system accuracy and efficiency are still ongoing. Currently, different modes of gantry and/or detector movement exist. The two main acquisition modes for acquiring DBT projection data are a continuous gantry motion mode and a step-and-shoot mode. In the continuous mode, the x-ray tube moves continuously while x-rays are generated in short pulses at specific angles around the breast. This has the potential for faster acquisition times compared with the step-and-shoot operation but incurs source motion blurring along the direction of the source motion. Source motion blurring has been shown to reduce technical image quality, as quantified by a number of metrics.1 – 5 In a step-and-shoot mode, the x-ray tube stops at each angular location to acquire a projection image. This eliminates source motion blur provided that the tube stops without inducing any mechanical instability but may result in a longer scan acquisition time and therefore an increased chance for patient motion artifacts.6